Comparison of CMM and Micro-CT Volumetric Analysis of Polyethylene Tibial Knee Inserts in Total Knee Replacement
المؤلفون المشاركون
Jiang, Wei
Jin, Zhongmin
Dai, Yuntian
Ji, Cuicui
Xiao, Huaxing
المصدر
International Journal of Polymer Science
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-7، 7ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-06-19
دولة النشر
مصر
عدد الصفحات
7
التخصصات الرئيسية
الملخص EN
Ultrahigh molecular weight polyethylene (UHMWPE) bearings are used widely in orthopaedic joint replacement as a prominent material for improving the longevity, which is dramatically influenced by wear of polyethylene.
Polyethylene tibial knee components from knee simulators under two different input conditions were analyzed using CMM and Micro-CT measurement techniques for volumetric loss with gravimetric measurement as reference.
Based on the coordinates and image slices obtained, the surface curve fitting and image digitization methodology were used for the creation of nominal “original” surface in the case of no prewear data provided.
The comparison results indicate that gravimetric remains the gold standard and the CMM measurement took less time and had better precision, accuracy, and repeatability compared to Micro-CT measurement technique.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Jiang, Wei& Ji, Cuicui& Xiao, Huaxing& Jin, Zhongmin& Dai, Yuntian. 2018. Comparison of CMM and Micro-CT Volumetric Analysis of Polyethylene Tibial Knee Inserts in Total Knee Replacement. International Journal of Polymer Science،Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1175075
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Jiang, Wei…[et al.]. Comparison of CMM and Micro-CT Volumetric Analysis of Polyethylene Tibial Knee Inserts in Total Knee Replacement. International Journal of Polymer Science No. 2018 (2018), pp.1-7.
https://search.emarefa.net/detail/BIM-1175075
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Jiang, Wei& Ji, Cuicui& Xiao, Huaxing& Jin, Zhongmin& Dai, Yuntian. Comparison of CMM and Micro-CT Volumetric Analysis of Polyethylene Tibial Knee Inserts in Total Knee Replacement. International Journal of Polymer Science. 2018. Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1175075
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1175075
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر