Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices
المؤلفون المشاركون
Kiraly, Laszlo
Stange, Jana
Kunert, Kathleen S.
Sel, Saadettin
المصدر
العدد
المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-8، 8ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2017-03-05
دولة النشر
مصر
عدد الصفحات
8
التخصصات الرئيسية
الملخص EN
Background.
To estimate repeatability and comparability of central corneal thickness (CCT) and keratometry measurements obtained by four different devices in healthy eyes.
Methods.
Fifty-five healthy eyes from 55 volunteers were enrolled in this study.
CCT (IOLMaster 700, Pentacam HR, and Cirrus HD-OCT) and keratometry readings (IOLMaster 700, Pentacam HR, and iDesign) were measured.
For statistical analysis, the corneal spherocylinder was converted into power vectors (J0, J45).
Repeatability was assessed by intraclass correlation coefficient (ICC).
Agreement of measurements between the devices was evaluated by the Bland-Altman method.
Results.
The analysis of repeatability of CCT data of IOLMaster 700, Pentacam HR, and Cirrus HD-OCT showed high ICCs (range 0.995 to 0.999).
The comparison of CCT measurements revealed statistically significant differences between Pentacam HR versus IOLMaster 700 (p<0.0001) and Pentacam HR versus Cirrus HD-OCT (p<0.0001), respectively.
There was no difference in CCT measurements between IOLMaster 700 and Cirrus HD-OCT (p=0.519).
The repeatability of keratometry readings (J0 and J45) of IOLMaster 700, Pentacam HR, and iDesign was also high with ICCs ranging from 0.974 to 0.999.
The Pentacam HR revealed significantly higher J0 in comparison to IOLMaster 700 (p=0.009) and iDesign (p=0.041); however, no significant difference was between IOLMaster 700 and iDesign (p=0.426).
Comparison of J45 showed no significant difference between IOLMaster 700, Pentacam HR, and iDesign.
These results were in accordance with Bland-Altman plots.
Conclusion.
In clinical practice, the devices analyzed should not be used interchangeably due to low agreement regarding CCT as well as keratometry readings.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Kiraly, Laszlo& Stange, Jana& Kunert, Kathleen S.& Sel, Saadettin. 2017. Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices. Journal of Ophthalmology،Vol. 2017, no. 2017, pp.1-8.
https://search.emarefa.net/detail/BIM-1185132
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Kiraly, Laszlo…[et al.]. Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices. Journal of Ophthalmology No. 2017 (2017), pp.1-8.
https://search.emarefa.net/detail/BIM-1185132
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Kiraly, Laszlo& Stange, Jana& Kunert, Kathleen S.& Sel, Saadettin. Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices. Journal of Ophthalmology. 2017. Vol. 2017, no. 2017, pp.1-8.
https://search.emarefa.net/detail/BIM-1185132
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1185132
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)