Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST
المؤلفون المشاركون
Vinciguerra, Riccardo
Vinciguerra, Paolo
Ambrósio, Renato
Koprowski, Robert
Lopes, Bernardo T.
Roberts, Cynthia J.
Elsheikh, Ahmed
Reisdorf, Sven
Berger, Stefanie
المصدر
العدد
المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-4، 4ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2017-06-06
دولة النشر
مصر
عدد الصفحات
4
التخصصات الرئيسية
الملخص EN
Purpose.
To assess the repeatability and reproducibility of dynamic corneal response parameters measured by the Corvis ST (Oculus, Wetzlar, Germany).
Methods.
One eye randomly selected from 32 healthy volunteers was examined by the Corvis ST.
Three different devices were used in an alternated random order for taking three measurements at each device in each subject.
Standard intraocular pressure (IOP), the biomechanical-compensated IOP (bIOP), and DCR parameters were evaluated.
The within-subject standard deviation (ζw) and coefficient of variation (CV) were assessed.
Results.
Regarding pressure indices, the ζw was below 1 mmHg for repeatability (0.98 for IOP and 0.89 for bIOP) and the CV was 6.6% for IOP and 6.1% for bIOP.
For reproducibility, the ζw was around 1 mmHg (1.12 for IOP and 1.05 for bIOP) and the CV was 7.6% for IOP and 7.1% for bIOP.
Most of DCR indices presented CV for repeatability below 4%.
For reproducibility, the CV of most of the indices were below 6%.
The deformation amplitude (DA) ratio in 1 mm and integrated radius were below 4% (1.2% and 3.8%, resp.).
Conclusions.
The Corvis ST showed good precision (repeatability and reproducibility) for IOP measurements and for DCR in healthy eyes.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Lopes, Bernardo T.& Roberts, Cynthia J.& Elsheikh, Ahmed& Vinciguerra, Riccardo& Vinciguerra, Paolo& Reisdorf, Sven…[et al.]. 2017. Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST. Journal of Ophthalmology،Vol. 2017, no. 2017, pp.1-4.
https://search.emarefa.net/detail/BIM-1185489
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Lopes, Bernardo T.…[et al.]. Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST. Journal of Ophthalmology No. 2017 (2017), pp.1-4.
https://search.emarefa.net/detail/BIM-1185489
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Lopes, Bernardo T.& Roberts, Cynthia J.& Elsheikh, Ahmed& Vinciguerra, Riccardo& Vinciguerra, Paolo& Reisdorf, Sven…[et al.]. Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST. Journal of Ophthalmology. 2017. Vol. 2017, no. 2017, pp.1-4.
https://search.emarefa.net/detail/BIM-1185489
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1185489
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)