Application of 14-MHz Ultrasonography with Tissue Harmonic Imaging to Determine Posterior Capsule Integrity in Traumatic Cataract
المؤلفون المشاركون
Yuan, Xiaoyong
Wu, Bin
Li, Qingyu
Liu, Yangchen
Wu, Di
Gao, Jianmin
Li, Yan
المصدر
العدد
المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2019-11-12
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Purpose.
To report the application of 14-MHz ultrasonography with tissue harmonic imaging (14 MHz + THI) to determine the integrity of the posterior capsule (PC) in traumatic cataract (TC).
Methods.
Patients with TC who were scheduled to undergo cataract extraction and whose PC could not be observed by slit lamp examination were included in the study.
The status of the PC was determined by 14 MHz + THI before cataract extraction and confirmed during surgery.
The results regarding PC integrity obtained from 14 MHz + THI and intraoperative direct observation were compared.
Result.
The study enrolled 52 eyes of 52 patients (49 men and 3 women), with a mean age of 42.15 years ± 11.23 (SD).
The nature of the trauma was blunt (3 eyes) or sharp (49 eyes).
The 14 MHz + THI method showed 21 PCs to be intact and 31 to have ruptured before cataract surgery.
During surgery, 23 PCs were observed to be intact, while 29 PCs were ruptured.
27 PCs were ruptured and 19 were intact, as determined by the two methods.
The 14 MHz + THI observations were consistent with the intraoperative observations of the PC (kappa = 0.764), with no significant difference between the two methods (P=0.687).
The sensitivity, specificity, and accuracy of 14 MHz + THI for observation of the PC were 93.10%, 82.60%, and 88.46%, respectively.
Conclusion.
The 14 MHz + THI method can accurately reveal the integrity of the PC in TC.
It has important clinical value in the selection of cataract surgery methods and the prediction of complications during TC surgery.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Wu, Bin& Li, Qingyu& Liu, Yangchen& Wu, Di& Gao, Jianmin& Li, Yan…[et al.]. 2019. Application of 14-MHz Ultrasonography with Tissue Harmonic Imaging to Determine Posterior Capsule Integrity in Traumatic Cataract. Journal of Ophthalmology،Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1185639
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Wu, Bin…[et al.]. Application of 14-MHz Ultrasonography with Tissue Harmonic Imaging to Determine Posterior Capsule Integrity in Traumatic Cataract. Journal of Ophthalmology No. 2019 (2019), pp.1-5.
https://search.emarefa.net/detail/BIM-1185639
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Wu, Bin& Li, Qingyu& Liu, Yangchen& Wu, Di& Gao, Jianmin& Li, Yan…[et al.]. Application of 14-MHz Ultrasonography with Tissue Harmonic Imaging to Determine Posterior Capsule Integrity in Traumatic Cataract. Journal of Ophthalmology. 2019. Vol. 2019, no. 2019, pp.1-5.
https://search.emarefa.net/detail/BIM-1185639
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1185639
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر