Comparative Analysis of Cutting Forces, Torques, and Vibration in Drilling of Bovine, Porcine, and Artificial Femur Bone with Considerations for Robot Effector Stiffness
المؤلفون المشاركون
Ibrahim, Dauda Sh.
Sharif, Umer
Orelaja, Oluseyi Adewale
Wang, Xingsong
Shen, Donghua
Zhao, Tianzheng
Afiz, Ishola A.
المصدر
Journal of Healthcare Engineering
العدد
المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-12، 12ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2020-10-20
دولة النشر
مصر
عدد الصفحات
12
التخصصات الرئيسية
الملخص EN
Bone drilling is known as one of the most sensitive milling processes in biomedical engineering field.
Fracture behavior of this cortical bone during drilling has attracted the attention of many researchers; however, there are still impending concerns such as necrosis, tool breakage, and microcracks due to high cutting forces, torques, and high vibration while drilling.
This paper presents a comparative analysis of the cutting forces, torques, and vibration resulted on different bone samples (bovine, porcine, and artificial femur) using a 6dof Robot arm effector with considerations of its stiffness effects.
Experiments were conducted on two spindle speeds of 1000 and 1500 rpm with a drill bit diameter of 2.5 mm and 6 mm depth of cut.
The results obtained from the specimens were processed and analyzed using MATLAB R2015b and Visio 2000 software; these results were then compared with a prior test using manual and conventional drilling methods.
The results obtained show that there is a significant drop in the average values of maximum drilling force for all the bone specimens when the spindle speed changes from 1000 rev/min to 1500 rev/min, with a drop from (20.07 to 12.34 N), approximately 23.85% for bovine, (11.25 to 8.14 N) with 16.03% for porcine, and (5.62 to 3.86 N) with 33.99% for artificial femur.
The maximum average values of torque also decrease from 41.2 to 24.2 N·mm (bovine), 37.0 to 21.6 N·mm (porcine), and 13.6 to 6.7 N·mm (artificial femur), respectively.
At an increase in the spindle speed, the vibration amplitude on all the bone samples also increases considerably.
The variation in drilling force, torque, and vibration in our result also confirm that the stiffness of the robot effector joint has negative effect on the bone precision during drilling process.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Orelaja, Oluseyi Adewale& Wang, Xingsong& Shen, Donghua& Ibrahim, Dauda Sh.& Zhao, Tianzheng& Sharif, Umer…[et al.]. 2020. Comparative Analysis of Cutting Forces, Torques, and Vibration in Drilling of Bovine, Porcine, and Artificial Femur Bone with Considerations for Robot Effector Stiffness. Journal of Healthcare Engineering،Vol. 2020, no. 2020, pp.1-12.
https://search.emarefa.net/detail/BIM-1186430
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Orelaja, Oluseyi Adewale…[et al.]. Comparative Analysis of Cutting Forces, Torques, and Vibration in Drilling of Bovine, Porcine, and Artificial Femur Bone with Considerations for Robot Effector Stiffness. Journal of Healthcare Engineering No. 2020 (2020), pp.1-12.
https://search.emarefa.net/detail/BIM-1186430
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Orelaja, Oluseyi Adewale& Wang, Xingsong& Shen, Donghua& Ibrahim, Dauda Sh.& Zhao, Tianzheng& Sharif, Umer…[et al.]. Comparative Analysis of Cutting Forces, Torques, and Vibration in Drilling of Bovine, Porcine, and Artificial Femur Bone with Considerations for Robot Effector Stiffness. Journal of Healthcare Engineering. 2020. Vol. 2020, no. 2020, pp.1-12.
https://search.emarefa.net/detail/BIM-1186430
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1186430
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر