The Difference in Repeatability of Automated Superficial Retinal Vessel Density according to the Measurement Area Using OCT Angiography

المؤلفون المشاركون

Lim, Hyung Bin
Kang, Tae Seen
Won, Yeo Kyoung
Kim, Jung Yeul

المصدر

Journal of Ophthalmology

العدد

المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-9، 9ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2020-04-17

دولة النشر

مصر

عدد الصفحات

9

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To evaluate the difference in the repeatability of automated superficial retinal vessel density and foveal avascular zone (FAZ) metrics according to the measurement area of optical coherence tomography angiography (OCTA).

Methods.

A total of 127 normal eyes from 127 healthy subjects were included.

Macular angiography images were acquired from all subjects using the Zeiss Cirrus 5000 with AngioPlex™ OCTA software.

Scans of 3 × 3 mm and 6 × 6 mm were each performed twice in a randomly arranged sequence.

Vessel density (VD), perfusion density (PD), and FAZ metrics of the superficial capillary plexus were calculated automatically for all scans, and the repeatabilities for both scan patterns were assessed based on intraclass correlation (ICC), coefficient of variation (CV), and coefficient of repeatability (CR) parameters.

The average measured values in the two scan patterns were also compared.

Results.

VD was significantly greater in the 3 × 3 mm scan than in the 6 × 6 mm scan according to all parameters, whereas PD was significantly less in the 3 × 3 mm scan than in the 6 × 6 mm scan.

The ICCs for VDs in the central fovea were 0.826 and 0.741 for the 3 × 3 and 6 × 6 mm scans, respectively, and the CVs were 8.00% and 12.75%.

For PDs, the ICCs were 0.839 and 0.762 and the CVs were 9.32% and 14.90%.

The FAZ metrics in the 3 × 3 mm scan showed good repeatability with an ICC >0.75 and a CV <10.0%.

However, all ICCs for the 6 × 6 mm scans were <0.75, and the CVs were all >10%.

Conclusions.

The 6 × 6 mm macular angiography scans resulted in lower repeatabilities than the 3 × 3 mm scans according to all OCTA parameters, particularly in the central fovea and FAZ metrics.

The 3 × 3 mm scan was more suitable than the 6 × 6 mm scan for analyzing macular microvascular density and FAZ metrics.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Lim, Hyung Bin& Kang, Tae Seen& Won, Yeo Kyoung& Kim, Jung Yeul. 2020. The Difference in Repeatability of Automated Superficial Retinal Vessel Density according to the Measurement Area Using OCT Angiography. Journal of Ophthalmology،Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1189487

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Lim, Hyung Bin…[et al.]. The Difference in Repeatability of Automated Superficial Retinal Vessel Density according to the Measurement Area Using OCT Angiography. Journal of Ophthalmology No. 2020 (2020), pp.1-9.
https://search.emarefa.net/detail/BIM-1189487

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Lim, Hyung Bin& Kang, Tae Seen& Won, Yeo Kyoung& Kim, Jung Yeul. The Difference in Repeatability of Automated Superficial Retinal Vessel Density according to the Measurement Area Using OCT Angiography. Journal of Ophthalmology. 2020. Vol. 2020, no. 2020, pp.1-9.
https://search.emarefa.net/detail/BIM-1189487

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1189487