Diagnostic Performance of a Novel Noninvasive Workup in the Setting of Dry Eye Disease

المؤلفون المشاركون

Pellegrini, Marco
Giannaccare, Giuseppe
Bernabei, Federico
Scorcia, Vincenzo
Vigo, Luca
Carones, Francesco

المصدر

Journal of Ophthalmology

العدد

المجلد 2020، العدد 2020 (31 ديسمبر/كانون الأول 2020)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2020-12-11

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To evaluate the diagnostic performance of a novel noninvasive automated workup employed for the diagnosis of dry eye disease (DED).

Methods.

One hundred patients with mild to moderate DED and 100 matched control subjects were enrolled in this cross-sectional study.

Ocular surface examinations were carried out by means of IDRA Plus (SBM Sistemi, Turin, Italy), which allows the automated evaluation of noninvasive breakup time (NIBUT), lipid layer thickness (LLT), tear meniscus height (TMH), infrared meibography for the measurement of meibomian gland loss (MGL), and blinking analysis.

Continuous variables were compared between patients with DED and controls by using the Mann–Whitney U test.

The area under the curve (AUC) of receiver operating characteristic curves was calculated.

The correlations between ocular surface parameters were evaluated with Pearson correlation analysis.

Results.

Patients with DED showed significantly lower values of NIBUT, LLT, and TMH compared to controls (6.9 ± 2.5 vs 10.4 ± 2.4 s, P < 0.001; 64.6 ± 20.3 vs 73.4 ± 21.9 nm, P = 0.003; 0.231 ± 0.115 vs 0.289 ± 0.164, P = 0.012, respectively).

Conversely, no significant differences were observed for MGL and blinking analysis (both P > 0.05).

NIBUT had the highest diagnostic power (AUC = 0.841, sensitivity = 0.89, and specificity = 0.69), followed by LLT (AUC = 0.621, sensitivity = 0.89, and specificity = 0.55), TMH (AUC = 0.606, sensitivity = 0.57, and specificity = 0.63), blink analysis (AUC = 0.533, sensitivity = 0.48, and specificity = 0.59), and MGL (AUC = 0.531, sensitivity = 0.54, and specificity = 0.48).

In patients with DED, NIBUT showed a significant correlation with TMH (R = 0.347, P = 0.002) and blinking analysis (R = 0.356, P < 0.001), while blinking analysis was negatively correlated with MGL (R = −0.315, P = 0.008).

Conclusions.

The automated noninvasive workup validated in this study may be a useful tool for reaching a noninvasive diagnosis of DED with a good performance, especially for NIBUT.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Vigo, Luca& Pellegrini, Marco& Bernabei, Federico& Carones, Francesco& Scorcia, Vincenzo& Giannaccare, Giuseppe. 2020. Diagnostic Performance of a Novel Noninvasive Workup in the Setting of Dry Eye Disease. Journal of Ophthalmology،Vol. 2020, no. 2020, pp.1-6.
https://search.emarefa.net/detail/BIM-1189498

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Vigo, Luca…[et al.]. Diagnostic Performance of a Novel Noninvasive Workup in the Setting of Dry Eye Disease. Journal of Ophthalmology No. 2020 (2020), pp.1-6.
https://search.emarefa.net/detail/BIM-1189498

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Vigo, Luca& Pellegrini, Marco& Bernabei, Federico& Carones, Francesco& Scorcia, Vincenzo& Giannaccare, Giuseppe. Diagnostic Performance of a Novel Noninvasive Workup in the Setting of Dry Eye Disease. Journal of Ophthalmology. 2020. Vol. 2020, no. 2020, pp.1-6.
https://search.emarefa.net/detail/BIM-1189498

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1189498