The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board

المؤلفون المشاركون

Chang, Ho
Kim, Kun-Yong
Lee, Win-Der
Cai, Yi-Fan
Chen, You-Jia

المصدر

Journal of Sensors

العدد

المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-8، 8ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2019-02-11

دولة النشر

مصر

عدد الصفحات

8

التخصصات الرئيسية

هندسة مدنية

الملخص EN

In this study, a novel gold-coated test strip for blood glucose measurement has been designed.

Such gold-coated test strip is feasible for mass production to achieve economies of scale.

Cyclic voltammetry was applied to test strips to undergo electrochemical reaction under a potential range of ±0.4 V.

Glucose oxidase (GOD) was added into K3[Fe(CN)6].

When glucose oxidase undergoes electrochemical reaction, the medium, K3[Fe(CN)6], will act as an electron acceptor, causing the electrodes on the test strip to generate a pair of clear anodic and reductive peaks.

The maximum of the anodic and reductive peaks can be used as reference to adjust the resistance of the blood glucose meter.

The experimental results show that by adjusting the resistance of the blood glucose meter, the accuracy of blood glucose meter reading can be tuned and blood glucose reading can be stabilized.

Therefore, when the resistance of the blood glucose meter is at 2.4 KΩ, the standard deviation (STD) and coefficient of variation (CV) of the test strip are lower than those of the test strips measured at resistances of 2.2 KΩ and 2.6 KΩ.

It has been proved in this study that adjusting the resistance of the blood glucose meter can optimize the chemical reaction on gold-coated test strips as well as its reading.

This method can also be applied to tune the accuracy of readings for test strips coated with other materials.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Kim, Kun-Yong& Chang, Ho& Lee, Win-Der& Cai, Yi-Fan& Chen, You-Jia. 2019. The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board. Journal of Sensors،Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191358

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Kim, Kun-Yong…[et al.]. The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board. Journal of Sensors No. 2019 (2019), pp.1-8.
https://search.emarefa.net/detail/BIM-1191358

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Kim, Kun-Yong& Chang, Ho& Lee, Win-Der& Cai, Yi-Fan& Chen, You-Jia. The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board. Journal of Sensors. 2019. Vol. 2019, no. 2019, pp.1-8.
https://search.emarefa.net/detail/BIM-1191358

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1191358