Design of Electrodes on Gold Test Strips for Enhanced Accuracy in Glucose Measurement
المؤلفون المشاركون
Chang, Ho
Kuo, Hsin-Yi
Cheng, Yuan-Hwei
Shaw, Jin-Siang
Lee, Rahnfong
المصدر
العدد
المجلد 2019، العدد 2019 (31 ديسمبر/كانون الأول 2019)، ص ص. 1-10، 10ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2019-10-15
دولة النشر
مصر
عدد الصفحات
10
التخصصات الرئيسية
الملخص EN
This paper reviews the practical process of designing the electrodes on electroless immersion gold-plated test strips for glucose measurement.
We have found that great care must be taken on determining the width and length of the working and reference electrodes, to yield the optimum performance on electroless immersion gold-plated glucose test strips for glucose measurement.
Preferably, a width of at least 0.6 mm and a length of at least 24 mm are recommended to prevent variance in electrode resistance during mass production of electroless immersion gold-plated glucose test strips, in order to enhance the accuracy of glucose measurement.
This paper also recommended that, if two different geometric sizes of test strips are going to be used on the same glucose meter, both test strips must use different lengths and widths on the working and reference electrodes to yield the same electrical resistance and hence similar electrochemical performance.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Kuo, Hsin-Yi& Cheng, Yuan-Hwei& Chang, Ho& Shaw, Jin-Siang& Lee, Rahnfong. 2019. Design of Electrodes on Gold Test Strips for Enhanced Accuracy in Glucose Measurement. Journal of Sensors،Vol. 2019, no. 2019, pp.1-10.
https://search.emarefa.net/detail/BIM-1191751
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Kuo, Hsin-Yi…[et al.]. Design of Electrodes on Gold Test Strips for Enhanced Accuracy in Glucose Measurement. Journal of Sensors No. 2019 (2019), pp.1-10.
https://search.emarefa.net/detail/BIM-1191751
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Kuo, Hsin-Yi& Cheng, Yuan-Hwei& Chang, Ho& Shaw, Jin-Siang& Lee, Rahnfong. Design of Electrodes on Gold Test Strips for Enhanced Accuracy in Glucose Measurement. Journal of Sensors. 2019. Vol. 2019, no. 2019, pp.1-10.
https://search.emarefa.net/detail/BIM-1191751
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1191751
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر