Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique
المؤلف
المصدر
العدد
المجلد 2017، العدد 2017 (31 ديسمبر/كانون الأول 2017)، ص ص. 1-7، 7ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2017-11-16
دولة النشر
مصر
عدد الصفحات
7
التخصصات الرئيسية
الملخص EN
Portable inhalers are divisible into those that deliver medication by patient triggering (pMDIs: a gentle slow inhalation) and those that use the patient’s inspiratory effort as the force for deaggregation and delivery (DPIs: a stronger deeper inspiratory effort).
Patient confusion and poor technique are commonplace.
The use of training tools has become standard practice, and unique amongst these is an inspiratory flow meter (In-Check) which is able to simulate the resistance characteristics of different inhalers and, thereby, guide the patient to the correct effort.
In-Check’s origins lie in the 1960s peak expiratory flow meters, the development of the Mini-Wright peak flow meter, and inspiratory flow assessment via the nose during the 1970s–1980s.
The current device (In-Check DIAL G16) is the third iteration of the original 1998 training tool, with detailed and ongoing assessments of all common inhaler resistances (including combination and breath-actuated inhaler types) summarised into resistance ranges that are preset within the device.
The device works by interpolating one of six ranges with the inspiratory effort.
Use of the tool has been shown to be contributory to significant improvements in asthma care and control, and it is being advocated for assessment and training in irreversible lung disease.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Sanders, Mark. 2017. Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique. Pulmonary Medicine،Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1197172
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Sanders, Mark. Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique. Pulmonary Medicine No. 2017 (2017), pp.1-7.
https://search.emarefa.net/detail/BIM-1197172
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Sanders, Mark. Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique. Pulmonary Medicine. 2017. Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1197172
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1197172
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر