Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters
المؤلفون المشاركون
Adams, Suzanne
Toroni, Bridget
Lele, Meenal
المصدر
العدد
المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2018-05-22
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Short peripheral catheters (SPC) are an existing conduit into many patients’ veins and line draws from SPC are a desired method of routine blood collection especially in difficult venous access patients.
The PIVO device facilitates blood collection through SPC and is being used clinically in a number of hospitals.
This study aimed to determine the appropriate wait time following a flush and the minimum waste volume required to obtain an undiluted blood sample when using the PIVO device and how that differed from current guidelines from SPC line draws.
A clinical study was conducted examining the analyte results of samples drawn with PIVO through a SPC at varying wait times following a saline flush.
Both an initial waste volume and a postwaste sample were compared to a venipuncture control.
The resulting samples showed no saline dilution as measured by sodium and creatinine results at all studied wait times.
These findings suggest that blood collections using the PIVO device can produce a clinically valid sample with a 30-second wait following a SPC flush and no waste volume prior to sample collection.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Adams, Suzanne& Toroni, Bridget& Lele, Meenal. 2018. Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters. Nursing Research and Practice،Vol. 2018, no. 2018, pp.1-6.
https://search.emarefa.net/detail/BIM-1210638
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Adams, Suzanne…[et al.]. Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters. Nursing Research and Practice No. 2018 (2018), pp.1-6.
https://search.emarefa.net/detail/BIM-1210638
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Adams, Suzanne& Toroni, Bridget& Lele, Meenal. Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters. Nursing Research and Practice. 2018. Vol. 2018, no. 2018, pp.1-6.
https://search.emarefa.net/detail/BIM-1210638
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-1210638
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر