Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician’s Dose from Scatter Radiation during Interventional Procedures

المؤلفون المشاركون

Sukupova, Lucie
Hlavacek, Ondrej
Vedlich, Daniel

المصدر

Radiology Research and Practice

العدد

المجلد 2018، العدد 2018 (31 ديسمبر/كانون الأول 2018)، ص ص. 1-7، 7ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2018-01-30

دولة النشر

مصر

عدد الصفحات

7

التخصصات الرئيسية

الطب البشري

الملخص EN

The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters.

Measurements were performed in different projections and with different positions of the ceiling-mounted shielding: without shielding; shielding closest to the patient; and shielding closest to the physician performing the procedure.

The protection provided by the shielding was assessed for cardiology when the femoral access is used and for radiology when the physician performs the procedure in the abdominal area.

The results show that the use of the ceiling-mounted shielding can decrease the dose from the scatter radiation by 95% at the position of the performing physician.

In cardiology, the impact is more pronounced when the left oblique projection is used.

In radiology, a large decrease was observed for right oblique projections, compared to cardiology.

The ceiling-mounted shielding should be placed as close to the physician as possible.

The idea of creating the largest radiation shadow by placing the radiation shielding as close to the patient as possible does not provide as effective radiation protection of the operator as it might be thought.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Sukupova, Lucie& Hlavacek, Ondrej& Vedlich, Daniel. 2018. Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician’s Dose from Scatter Radiation during Interventional Procedures. Radiology Research and Practice،Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1212763

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Sukupova, Lucie…[et al.]. Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician’s Dose from Scatter Radiation during Interventional Procedures. Radiology Research and Practice No. 2018 (2018), pp.1-7.
https://search.emarefa.net/detail/BIM-1212763

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Sukupova, Lucie& Hlavacek, Ondrej& Vedlich, Daniel. Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician’s Dose from Scatter Radiation during Interventional Procedures. Radiology Research and Practice. 2018. Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1212763

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-1212763