Consistency of CT KUB radiation dose and exposure parameters in King Abdulaziz Medical City in Jeddah : quality assessment

العناوين الأخرى

تناسق الجرعة الإشعاعية للتصوير المقطعي للكلى و الحالب و المثانة في مدينة الملك عبد العزيز الطبية بجدة : تقييم الجودة

المصدر

Journal of King Abdulaziz University : Engineering Sciences

العدد

المجلد 30، العدد 2 (31 ديسمبر/كانون الأول 2019)، ص ص. 37-42، 6ص.

الناشر

جامعة الملك عبد العزيز مركز النشر العلمي

تاريخ النشر

2019-12-31

دولة النشر

السعودية

عدد الصفحات

6

التخصصات الرئيسية

العلوم التربوية

الملخص EN

Computed Tomography (CT) is a type of Medical Imaging that uses X-ray to generate cross-sectional images of the body with an effective dose that is higher than the dose of conventional radiography.

For instance, the radiation dose for Kidney, Ureter, and Bladder (KUB) in conventional radiography is 0.7mSv, whereas, in the CT scan is 10mSv.

Therefore, this study aimed to evaluate the patient radiation dose delivered in unenhanced CT-KUB examinations using two CT machines in compliance with the American College of Radiology (ACR) standards.

This was a retrospective, cross-sectional study of all patients who had unenhanced CT-KUB examinations in King Abdulaziz Medical City in Jeddah between January 1st and June 30th, 2018.

A total of 264 patients were included in the study.

The mean CTDIvol, DLP, Time (s), pitch and effective dose were significantly different in the CT machines with a p-value between 0.004-<0.001.

Both CT machines were within the ACR standards range of 10mSv.

There was a positive strong correlation between the BMI (29.6 kg/m2) and the effective dose with p-value=<001.

The study further revealed that radiation doses delivered by the two CT scanners are compliant with the ACR standards.-

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

al-Mutayri, Rahaf H.. 2019. Consistency of CT KUB radiation dose and exposure parameters in King Abdulaziz Medical City in Jeddah : quality assessment. Journal of King Abdulaziz University : Engineering Sciences،Vol. 30, no. 2, pp.37-42.
https://search.emarefa.net/detail/BIM-1243803

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

al-Mutayri, Rahaf H.. Consistency of CT KUB radiation dose and exposure parameters in King Abdulaziz Medical City in Jeddah : quality assessment. Journal of King Abdulaziz University : Engineering Sciences Vol. 30, no. 2 (2019), pp.37-42.
https://search.emarefa.net/detail/BIM-1243803

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

al-Mutayri, Rahaf H.. Consistency of CT KUB radiation dose and exposure parameters in King Abdulaziz Medical City in Jeddah : quality assessment. Journal of King Abdulaziz University : Engineering Sciences. 2019. Vol. 30, no. 2, pp.37-42.
https://search.emarefa.net/detail/BIM-1243803

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

رقم السجل

BIM-1243803