Endotracheal tube cuff pressure monitoring in intensive care units
المؤلفون المشاركون
Nekooeian, A. A.
Zand, Farid
Rohani, M.
المصدر
Iranian Red Crescent Medical Journal
العدد
المجلد 10، العدد 3 (31 يوليو/تموز 2008)، ص ص. 220-224، 5ص.
الناشر
تاريخ النشر
2008-07-31
دولة النشر
الإمارات العربية المتحدة
عدد الصفحات
5
التخصصات الرئيسية
الموضوعات
الملخص EN
Background : improper inflation of endotracheal tube cuff is associated with various complications.
Recently, there has been high referral of patients with post-intubation complications to the clinics affiliated to Shiraz University of Medical Sciences.
Herein, we examined the practice of cuff pressure monitoring at 9 adult intensive care units in 3 university-affiliated hospitals.
Methods : the present study was performed on 57 tracheal intubated patients, whose cuff pressures were measured using either a cuff pressure gauge on 12 patients (Group 1) or manual palpation of the pilot balloon on 45 cases (Group 2) during summer 2004.
The intracuff pressure (P1), intracuff volume (V1), and pressure exerted by the cuff against the tracheal mucosa (ΔP) were determined, analyzed and compared with the recommended values.
Results : the P1 in Group 1 was 35.3 ± 32.8 SD cmH2O, with 33.3 % of the patients having values below 20 cmH2O.
In group two, P1 was 88.8 ± 27.1 cmH2O, all of which being above 40 cmH2O.
The ΔP for the Groups 1 and 2 were 16.3 ± 11.1SD cm H2O and 35.5 ±15.6 cm H2O, respectively.
Conclusion : our findings indicated that manual palpation of balloon cuff to monitor cuff pressure was not an accurate procedure.
They also suggested that measures to increase the knowledge and skills of ICU staff and posting of cuff pressure monitoring protocols might help improve the practice.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zand, Farid& Nekooeian, A. A.& Rohani, M.. 2008. Endotracheal tube cuff pressure monitoring in intensive care units. Iranian Red Crescent Medical Journal،Vol. 10, no. 3, pp.220-224.
https://search.emarefa.net/detail/BIM-28209
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zand, Farid…[et al.]. Endotracheal tube cuff pressure monitoring in intensive care units. Iranian Red Crescent Medical Journal Vol. 10, no. 3 (Jul. 2008), pp.220-224.
https://search.emarefa.net/detail/BIM-28209
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zand, Farid& Nekooeian, A. A.& Rohani, M.. Endotracheal tube cuff pressure monitoring in intensive care units. Iranian Red Crescent Medical Journal. 2008. Vol. 10, no. 3, pp.220-224.
https://search.emarefa.net/detail/BIM-28209
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references : p. 223-224
رقم السجل
BIM-28209
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)