![](/images/graphics-bg.png)
Linear image distortion of different patients' head position using dimax direct digital panoramic x-ray system
المؤلف
المصدر
Journal of Baghdad College of Dentistry
العدد
المجلد 19، العدد 1 (31 ديسمبر/كانون الأول 2007)، ص ص. 65-68، 4ص.
الناشر
تاريخ النشر
2007-12-31
دولة النشر
العراق
عدد الصفحات
4
التخصصات الرئيسية
الموضوعات
الملخص EN
Background: The aim of this study is to show the effect of improper patient head positioning on the distortion amount of x-ray images.
Material and method: Distortion amount is measured using a special ball phantom which contain 23 balls to check the patient positioning mechanisms using Planmeca 2002 CC Praline Digital X-ray unit by adjustment the position to 00 both x- and y- directions, then linear movement was done by displacement of the ball phantom +5mn anteriorly and-5mm posteriorly.
Exposure to X-rays was done for each position.
The three images were seen on computer screen, balls and space in between dimensions were measured using Dimax software.
Results: The results revealed that at zero position, balls appear round, all with same size, every spaced with constant magnification factors.
When the ball phantom displaced, different distortion amount in the balls will take place especially in the anterior region.
Conclusion: Exact head positioning must be taken in consideration to ensure that wanted area is included within the zone of sharpness.
Key words: Radiography distortion digital panoramic.
(J Bagh Coll Dentistry 2007; 19(1)65-68)
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
al-Naqib, Lamya H.. 2007. Linear image distortion of different patients' head position using dimax direct digital panoramic x-ray system. Journal of Baghdad College of Dentistry،Vol. 19, no. 1, pp.65-68.
https://search.emarefa.net/detail/BIM-288499
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
al-Naqib, Lamya H.. Linear image distortion of different patients' head position using dimax direct digital panoramic x-ray system. Journal of Baghdad College of Dentistry Vol. 19, no. 1 (2007), pp.65-68.
https://search.emarefa.net/detail/BIM-288499
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
al-Naqib, Lamya H.. Linear image distortion of different patients' head position using dimax direct digital panoramic x-ray system. Journal of Baghdad College of Dentistry. 2007. Vol. 19, no. 1, pp.65-68.
https://search.emarefa.net/detail/BIM-288499
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references : p. 68
رقم السجل
BIM-288499
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)