Background: Accurate working length determination is a very important aspect of root canal therapy.
There have been many methods developed to estimate the length of a canal.
One of the most accurate methods of determining the length of a canal is the use of an electronic apex locating device.
Materials and Methods: In this in vivo study twenty lower single rooted teeth were used, group A which involved ten vital teeth (extracted for orthodontic treatment) and group B which involved also ten non-vital teeth .The two groups were tested by the use of the Root Zc apex locator to estimate the working length.
Results: From the data obtained there was no difference in the clinical efficiency of the electronic apex locator in estimating the working length whether the tooth was vital or not.
Conclusion: The Root Zx apex locator was extremely accurate in determining the canal length when used in vital and non-vital teeth.
Keywords: Electronic apex locators, Root Zx apex locator.
(J Bagh Coll Dentistry 2008; 20(2):50-52)
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Hani, Firas I.& Abd al-Baqi, Muhammad A.. 2008. An in vivo study to evaluate the clinical accuracy of Root Zx apex locator in vital and necrotic teeth. Journal of Baghdad College of Dentistry،Vol. 20, no. 2, pp.50-52.
https://search.emarefa.net/detail/BIM-288812
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Hani, Firas I.& Abd al-Baqi, Muhammad A.. An in vivo study to evaluate the clinical accuracy of Root Zx apex locator in vital and necrotic teeth. Journal of Baghdad College of Dentistry Vol. 20, no. 2 (2008), pp.50-52.
https://search.emarefa.net/detail/BIM-288812
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Hani, Firas I.& Abd al-Baqi, Muhammad A.. An in vivo study to evaluate the clinical accuracy of Root Zx apex locator in vital and necrotic teeth. Journal of Baghdad College of Dentistry. 2008. Vol. 20, no. 2, pp.50-52.
https://search.emarefa.net/detail/BIM-288812
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references : p. 52
رقم السجل
BIM-288812
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر