Purpose.
To test different peripheral arterial stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation and image noise in dual-source multidetector row CT (DSCT) in vitro.
Methods and Materials.
22 stents (nitinol, steel, cobalt-alloy, tantalum, platinum alloy) were examined in a vessel phantom.
All stents were imaged in axial orientation with standard parameters.
Image reconstructions were obtained with four different convolution kernels.
To evaluate visualization characteristics of the stent, the lumen diameter, intraluminal density and noise were measured.
Results.
The mean percentage of the visible stent lumen diameter from the nominal stent diameter was 74.5% ± 5.7 for the medium-sharp kernel, 72.8% ± 6.4 for the medium, 70.8% ± 6.4 for the medium-smooth and 67.6% ± 6.6 for the smooth kernel.
Mean values of lumen attenuation were 299.7HU ± 127 (medium-sharp), 273.9HU ± 68 (medium), 270.7HU ± 53 (medium-smooth) and 265.8HU ± 43.
Mean image noise was: 54.6 ± 6.3, 20.5 ± 1.7, 16.3 ± 1.7, 14.0 ± 2 respectively.
Conclusion.
Visible stent lumen diameter varies depending on stent type and scan parameters.
Lumen diameter visibility increases with the sharpness of the reconstruction kernel.
Smoother kernels provide more realistic density measurements inside the stent lumen and less image noise.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Köhler, Michael& Burg, Matthias C.& Bunck, Alexander C.& Heindel, Walter& Seifarth, Harald& Maintz, David. 2011. Dual-Source CT Angiography of Peripheral Arterial Stents : In Vitro Evaluation of 22 Different Stent Types. Radiology Research and Practice،Vol. 2011, no. 2011, pp.1-7.
https://search.emarefa.net/detail/BIM-446622
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Köhler, Michael…[et al.]. Dual-Source CT Angiography of Peripheral Arterial Stents : In Vitro Evaluation of 22 Different Stent Types. Radiology Research and Practice No. 2011 (2011), pp.1-7.
https://search.emarefa.net/detail/BIM-446622
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Köhler, Michael& Burg, Matthias C.& Bunck, Alexander C.& Heindel, Walter& Seifarth, Harald& Maintz, David. Dual-Source CT Angiography of Peripheral Arterial Stents : In Vitro Evaluation of 22 Different Stent Types. Radiology Research and Practice. 2011. Vol. 2011, no. 2011, pp.1-7.
https://search.emarefa.net/detail/BIM-446622
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-446622
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)