A Fluid Mechanical Interpretation of Hysteresis in Rhinomanometry

المؤلفون المشاركون

Peters, F.
Groß, T. F.

المصدر

ISRN Otolaryngology

العدد

المجلد 2011، العدد 2011 (31 ديسمبر/كانون الأول 2011)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-09-21

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الأمراض
الطب البشري

الملخص EN

A hysteresis effect in the pressure/flow rate relationship of nasal breathing has frequently been observed in clinical tests and in lab investigations.

Explanations that have been given in the literature are missing a fluid mechanic storage effect coming into play in reciprocating flows.

This effect depends primarily on the way the rhinomanometric measurements are set up and not so much on the nose flow itself.

This is to be shown by calculations and experiments.

The experiments are carried out with orifices because they can represent nose flow and are often implemented in rhinomanometric equipment as flow gauges.

To mimic reality also a 1 : 1 nose model is used.

It is shown where the hysteresis comes from and what the key parameters for its prediction are.

With these results hysteresis in nasal breathing appears in a new light.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Groß, T. F.& Peters, F.. 2011. A Fluid Mechanical Interpretation of Hysteresis in Rhinomanometry. ISRN Otolaryngology،Vol. 2011, no. 2011, pp.1-6.
https://search.emarefa.net/detail/BIM-447715

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Groß, T. F.& Peters, F.. A Fluid Mechanical Interpretation of Hysteresis in Rhinomanometry. ISRN Otolaryngology No. 2011 (2011), pp.1-6.
https://search.emarefa.net/detail/BIM-447715

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Groß, T. F.& Peters, F.. A Fluid Mechanical Interpretation of Hysteresis in Rhinomanometry. ISRN Otolaryngology. 2011. Vol. 2011, no. 2011, pp.1-6.
https://search.emarefa.net/detail/BIM-447715

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-447715