Study on Microchannel Design and Burst Frequency Detection for Centrifugal Microfluidic System
المؤلفون المشاركون
Chang, Yaw-Jen
Chen, Shia-Chung
Hsu, Cheng-Li
المصدر
Advances in Materials Science and Engineering
العدد
المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2013-02-17
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
العلوم الهندسية و تكنولوجيا المعلومات
الملخص EN
A centrifugal microfluidic system has been developed in this study, enabling the control and measurement of the burst frequency in order to manipulate the liquid.
The radial microfluid chips with different microchannel dimensions were designed for simulation analyses and experimental verifications.
The microfluidic flow in the microchannel was analyzed using software CFDRC, providing an accurate result compared with that from experiment.
The results show that the design of the overflow microchannel can correctly keep the liquid volume with error as low as 5%.
For mercurochrome, the burst frequency has an inverse proportion to the channel width, and the simulation results agree with the experimental results.
For oil, however, the experimental and simulation results indicate that the relationship between the burst frequency and channel width is not obvious due to oil properties.
Since the simulation approach can provide an accurate prediction of flow behavior in the microchannel, the design of radial microfluid chip and the control of burst frequency can be achieved effectively.
A practical application to design the centrifugal microfluidic disc for blood typing test was also carried out in this study.
The centrifugal microfluidic system can successfully control the spinning speed to achieve the result of adding reagents in a specific sequence.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Chang, Yaw-Jen& Chen, Shia-Chung& Hsu, Cheng-Li. 2013. Study on Microchannel Design and Burst Frequency Detection for Centrifugal Microfluidic System. Advances in Materials Science and Engineering،Vol. 2013, no. 2013, pp.1-9.
https://search.emarefa.net/detail/BIM-448648
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Chang, Yaw-Jen…[et al.]. Study on Microchannel Design and Burst Frequency Detection for Centrifugal Microfluidic System. Advances in Materials Science and Engineering No. 2013 (2013), pp.1-9.
https://search.emarefa.net/detail/BIM-448648
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Chang, Yaw-Jen& Chen, Shia-Chung& Hsu, Cheng-Li. Study on Microchannel Design and Burst Frequency Detection for Centrifugal Microfluidic System. Advances in Materials Science and Engineering. 2013. Vol. 2013, no. 2013, pp.1-9.
https://search.emarefa.net/detail/BIM-448648
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-448648
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر