A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature

المؤلفون المشاركون

Lim, Justin
Morley, David
Starks, Ian

المصدر

Case Reports in Orthopedics

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-4، 4ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2012-12-09

دولة النشر

مصر

عدد الصفحات

4

التخصصات الرئيسية

الطب البشري

الملخص EN

We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation.

This occurred at the head-neck junction.

Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion.

The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Morley, David& Starks, Ian& Lim, Justin. 2012. A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature. Case Reports in Orthopedics،Vol. 2012, no. 2012, pp.1-4.
https://search.emarefa.net/detail/BIM-450499

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Morley, David…[et al.]. A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature. Case Reports in Orthopedics No. 2012 (2012), pp.1-4.
https://search.emarefa.net/detail/BIM-450499

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Morley, David& Starks, Ian& Lim, Justin. A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature. Case Reports in Orthopedics. 2012. Vol. 2012, no. 2012, pp.1-4.
https://search.emarefa.net/detail/BIM-450499

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-450499