Breakage of a Third Generation Gamma Nail : A Case Report and Review of the Literature

المؤلفون المشاركون

Lee, Sang Yang
Iwakura, Takashi
Sakai, Yoshitada
Kuroda, Ryosuke
Kurosaka, Masahiro
Nishida, Kotaro
Niikura, Takahiro

المصدر

Case Reports in Orthopedics

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2013-05-20

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

The use of intramedullary nails to treat trochanteric fractures of the femur has increased with the increasing size of the elderly population.

The third generation Gamma nail is currently one of the most popular devices for the treatment of trochanteric fractures.

Nail breakage is a rare complication, possibly resulting from fatigue fracture of the implant.

We present the first reported case of breakage of a third generation Gamma nail that was not used to treat a pathological fracture.

An 83-year-old woman with an unstable trochanteric fracture of the femur was treated using a third generation Gamma nail.

She was referred to our hospital 14 months postoperatively with nail breakage at the opening for the lag screw.

The breakage was secondary to nonunion, which was thought to be mainly due to insufficient reduction of the fracture.

The broken nail was removed, and the patient underwent cemented bipolar hemiarthroplasty.

At followup 18 months later, she was mobile with a walker and asymptomatic with no complications.

This case shows that inadequate operation such as insufficient reduction of the trochanteric fracture may result in nonunion and implant breakage, even when using a high-strength, well-designed implant.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Iwakura, Takashi& Niikura, Takahiro& Lee, Sang Yang& Sakai, Yoshitada& Nishida, Kotaro& Kuroda, Ryosuke…[et al.]. 2013. Breakage of a Third Generation Gamma Nail : A Case Report and Review of the Literature. Case Reports in Orthopedics،Vol. 2013, no. 2013, pp.1-5.
https://search.emarefa.net/detail/BIM-451647

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Iwakura, Takashi…[et al.]. Breakage of a Third Generation Gamma Nail : A Case Report and Review of the Literature. Case Reports in Orthopedics No. 2013 (2013), pp.1-5.
https://search.emarefa.net/detail/BIM-451647

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Iwakura, Takashi& Niikura, Takahiro& Lee, Sang Yang& Sakai, Yoshitada& Nishida, Kotaro& Kuroda, Ryosuke…[et al.]. Breakage of a Third Generation Gamma Nail : A Case Report and Review of the Literature. Case Reports in Orthopedics. 2013. Vol. 2013, no. 2013, pp.1-5.
https://search.emarefa.net/detail/BIM-451647

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-451647