Purpose.

To evaluate the clinical applicability of a telemonitoring system: telemetric system for collection and distant surveillance of medical information (TEMEO).

Methods.

We evaluated 60 patients, applying simultaneously standard Holter ECG and telemonitoring.

Two different comparisons were performed: (1) TEMEO ECG with standard 12-lead ECG; (2) TEMEO Holter with standard ECG Holter.

Results.

We found a very high coincidence rate (99.3%) between TEMEO derived ECGs and standard ECGs.

Intraclass correlation coefficient analysis revealed high and significant correlation coefficients regarding average, maximal, and minimal heart rate, % of time in tachycardia, single supraventricular ectopic beats (SVEB), and single and couplets of ventricular ectopic beats (VEB) between Holter ECG and TEMEO derived parameters.

Couplets of SVEB were recorded as different by the two monitoring systems, however, with a borderline statistical significance.

Conclusions.

TEMEO derived ECGs have a very high coincidence rate with standard ECGs.

TEMEO patient monitoring provides results that are similar to those derived from a standard Holter ECG.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Mateev, Hristo& Simova, Iana& Katova, Tzvetana& Dimitrov, Nikolay. 2012. Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System. ISRN Cardiology،Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-453338

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Mateev, Hristo…[et al.]. Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System. ISRN Cardiology No. 2012 (2012), pp.1-8.
https://search.emarefa.net/detail/BIM-453338

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Mateev, Hristo& Simova, Iana& Katova, Tzvetana& Dimitrov, Nikolay. Clinical Evaluation of a Mobile Heart Rhythm Telemonitoring System. ISRN Cardiology. 2012. Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-453338

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-453338