Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty

المؤلفون المشاركون

Higashi, Hidetaka
Watanabe, Hitoshi
Kubosawa, Hitoshi
Sakamoto, Masaaki

المصدر

Case Reports in Orthopedics

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-4، 4ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-03-04

دولة النشر

مصر

عدد الصفحات

4

التخصصات الرئيسية

الطب البشري

الملخص EN

A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered.

However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion.

Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed.

Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself.

Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Sakamoto, Masaaki& Watanabe, Hitoshi& Higashi, Hidetaka& Kubosawa, Hitoshi. 2014. Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty. Case Reports in Orthopedics،Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-454743

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Sakamoto, Masaaki…[et al.]. Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty. Case Reports in Orthopedics No. 2014 (2014), pp.1-4.
https://search.emarefa.net/detail/BIM-454743

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Sakamoto, Masaaki& Watanabe, Hitoshi& Higashi, Hidetaka& Kubosawa, Hitoshi. Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty. Case Reports in Orthopedics. 2014. Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-454743

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-454743