الملخص EN
A simple and sensitive high performance liquid chromatography-tandem mass spectrometry (HPLC-ESI-MS-MS) method was developed and validated for the quantification of creatinine in human urine.
The analysis was carried out on an Agilent Zorbax Eclipse XDB-C18 column (2.1×150 mm, 3.5 μm).
The mobile phase was 0.1% formic acid in water and 0.1% formic acid in acetonitrile (50/50, v/v).
Linear calibration curves were obtained in the concentration range of 1–2000.0 ng/mL, with a lower limit of quantification of 0.99 ng/mL.
The intra- and interday precision (RSD) values were below 3%.
The method was successfully applied to a bioequivalence study of creatinine in Chinese smokers and nonsmokers.
The total cotinine in 24 h urine and cotinine : creatinine ratio were also positively associated (Pearson R=0.942, P<0.0001).
However, cotinine : creatinine ratio varied significantly across smoking groups for the difference of individual.
24 h urinary cotinine was more appropriate for expressing correlation with tar than cotinine : creatinine ratio.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Hou, Hongwei& Xiong, Wei& Zhang, Xiaotao& Song, Dongkui& Tang, Gangling& Hu, Qingyuan. 2012. LC-MS-MS Measurements of Urinary Creatinine and the Application of Creatinine Normalization Technique on Cotinine in Smokers’ 24 Hour Urine. Journal of Analytical Methods in Chemistry،Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-456905
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Hou, Hongwei…[et al.]. LC-MS-MS Measurements of Urinary Creatinine and the Application of Creatinine Normalization Technique on Cotinine in Smokers’ 24 Hour Urine. Journal of Analytical Methods in Chemistry No. 2012 (2012), pp.1-8.
https://search.emarefa.net/detail/BIM-456905
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Hou, Hongwei& Xiong, Wei& Zhang, Xiaotao& Song, Dongkui& Tang, Gangling& Hu, Qingyuan. LC-MS-MS Measurements of Urinary Creatinine and the Application of Creatinine Normalization Technique on Cotinine in Smokers’ 24 Hour Urine. Journal of Analytical Methods in Chemistry. 2012. Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-456905
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-456905
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر