![](/images/graphics-bg.png)
Modiolus-Hugging Intracochlear Electrode Array with Shape Memory Alloy
المؤلفون المشاركون
Min, Kyou Sik
Park, Se-Ik
Jun, Sang Beom
Lim, Yoon Seob
Kim, Sung June
المصدر
Computational and Mathematical Methods in Medicine
العدد
المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2013-05-09
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
In the cochlear implant system, the distance between spiral ganglia and the electrodes within the volume of the scala tympani cavity significantly affects the efficiency of the electrical stimulation in terms of the threshold current level and spatial selectivity.
Because the spiral ganglia are situated inside the modiolus, the central axis of the cochlea, it is desirable that the electrode array hugs the modiolus to minimize the distance between the electrodes and the ganglia.
In the present study, we propose a shape-memory-alloy-(SMA-) embedded intracochlear electrode which gives a straight electrode a curved modiolus-hugging shape using the restoration force of the SMA as triggered by resistive heating after insertion into the cochlea.
An eight-channel ball-type electrode array is fabricated with an embedded titanium-nickel SMA backbone wire.
It is demonstrated that the electrode array changes its shape in a transparent plastic human cochlear model.
To verify the safe insertion of the electrode array into the human cochlea, the contact pressures during insertion at the electrode tip and the contact pressures over the electrode length after insertion were calculated using a 3D finite element analysis.
The results indicate that the SMA-embedded electrode is functionally and mechanically feasible for clinical applications.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Min, Kyou Sik& Jun, Sang Beom& Lim, Yoon Seob& Park, Se-Ik& Kim, Sung June. 2013. Modiolus-Hugging Intracochlear Electrode Array with Shape Memory Alloy. Computational and Mathematical Methods in Medicine،Vol. 2013, no. 2013, pp.1-9.
https://search.emarefa.net/detail/BIM-457446
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Min, Kyou Sik…[et al.]. Modiolus-Hugging Intracochlear Electrode Array with Shape Memory Alloy. Computational and Mathematical Methods in Medicine No. 2013 (2013), pp.1-9.
https://search.emarefa.net/detail/BIM-457446
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Min, Kyou Sik& Jun, Sang Beom& Lim, Yoon Seob& Park, Se-Ik& Kim, Sung June. Modiolus-Hugging Intracochlear Electrode Array with Shape Memory Alloy. Computational and Mathematical Methods in Medicine. 2013. Vol. 2013, no. 2013, pp.1-9.
https://search.emarefa.net/detail/BIM-457446
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-457446
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)