Comparative Evaluation of Osseointegrated Dental Implants Based on Platform-Switching Concept : Influence of Diameter, Length, Thread Shape, and In-Bone Positioning Depth on Stress-Based Performance

المؤلفون المشاركون

Sannino, Gianpaolo
Vairo, Giuseppe

المصدر

Computational and Mathematical Methods in Medicine

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-15، 15ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2013-06-19

دولة النشر

مصر

عدد الصفحات

15

التخصصات الرئيسية

الطب البشري

الملخص EN

This study aimed to investigate the influence of implant design (in terms of diameter, length, and thread shape), in-bone positioning depth, and bone posthealing crestal morphology on load transfer mechanisms of osseointegrated dental implants based on platform-switching concept.

In order to perform an effective multiparametric comparative analysis, 11 implants different in dimensions and in thread features were analyzed by a linearly elastic 3-dimensional finite element approach, under a static load.

Implant models were integrated with the detailed model of a maxillary premolar bone segment.

Different implant in-bone positioning levels were modeled, considering also different posthealing crestal bone morphologies.

Bone overloading risk was quantified by introducing proper local stress measures, highlighting that implant diameter is a more effective design parameter than the implant length, as well as that thread shape and thread details can significantly affect stresses at peri-implant bone, especially for short implants.

Numerical simulations revealed that the optimal in-bone positioning depth results from the balance of 2 counteracting effects: cratering phenomena and bone apposition induced by platform-switching configuration.

Proposed results contribute to identify the mutual influence of a number of factors affecting the bone-implant loading transfer mechanisms, furnishing useful insights and indications for choosing and/or designing threaded osseointegrated implants.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Vairo, Giuseppe& Sannino, Gianpaolo. 2013. Comparative Evaluation of Osseointegrated Dental Implants Based on Platform-Switching Concept : Influence of Diameter, Length, Thread Shape, and In-Bone Positioning Depth on Stress-Based Performance. Computational and Mathematical Methods in Medicine،Vol. 2013, no. 2013, pp.1-15.
https://search.emarefa.net/detail/BIM-457447

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Vairo, Giuseppe& Sannino, Gianpaolo. Comparative Evaluation of Osseointegrated Dental Implants Based on Platform-Switching Concept : Influence of Diameter, Length, Thread Shape, and In-Bone Positioning Depth on Stress-Based Performance. Computational and Mathematical Methods in Medicine No. 2013 (2013), pp.1-15.
https://search.emarefa.net/detail/BIM-457447

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Vairo, Giuseppe& Sannino, Gianpaolo. Comparative Evaluation of Osseointegrated Dental Implants Based on Platform-Switching Concept : Influence of Diameter, Length, Thread Shape, and In-Bone Positioning Depth on Stress-Based Performance. Computational and Mathematical Methods in Medicine. 2013. Vol. 2013, no. 2013, pp.1-15.
https://search.emarefa.net/detail/BIM-457447

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-457447