Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO

المؤلفون المشاركون

Stemmler, Hans-Joachim
Cnossen, Jitske
Geiger, Sandra
Brettner, Florian
Horster, Sophia
Nickel, Thomas
Parhofer, Klaus G.
Strecker, Nina
Hausmann, Andreas

المصدر

Critical Care Research and Practice

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-11-29

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

USCOM is an ultrasound-based method which has been accepted for noninvasive hemodynamic monitoring in various clinical conditions (USCOM, Ultrasonic cardiac output monitoring).

The present study aimed at comparing the accuracy of the USCOM device with that of the thermodilution technique in patients with septicemia.

We conducted a prospective observational study in a medical but noncardiological ICU of a university hospital.

Septic adult patients (median age 55 years, median SAPS-II-Score 43 points) on mechanical ventilation and catecholamine support were monitored with USCOM and PiCCO (n=70).

Seventy paired left-sided CO measurements (transaortic access = COUS-A) were obtained.

The mean COUS-A were 6.55 l/min (±2.19) versus COPiCCO 6.5 l/min (±2.18).

The correlation coefficient was r=0.89.

Comparison by Bland-Altman analysis revealed a bias of −0.36 l/min (±0.99 l/min) leading to a mean percentage error of 29%.

USCOM is a feasible and rapid method to evaluate CO in septic patients.

USCOM does reliably represent CO values as compared to the reference technique based on thermodilution (PiCCO).

It seems to be appropriate in situations where CO measurements are most pertinent to patient management.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Horster, Sophia& Stemmler, Hans-Joachim& Strecker, Nina& Brettner, Florian& Hausmann, Andreas& Cnossen, Jitske…[et al.]. 2011. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-459125

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Horster, Sophia…[et al.]. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-459125

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Horster, Sophia& Stemmler, Hans-Joachim& Strecker, Nina& Brettner, Florian& Hausmann, Andreas& Cnossen, Jitske…[et al.]. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice. 2011. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-459125

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-459125