Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO
المؤلفون المشاركون
Stemmler, Hans-Joachim
Cnossen, Jitske
Geiger, Sandra
Brettner, Florian
Horster, Sophia
Nickel, Thomas
Parhofer, Klaus G.
Strecker, Nina
Hausmann, Andreas
المصدر
Critical Care Research and Practice
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2011-11-29
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
USCOM is an ultrasound-based method which has been accepted for noninvasive hemodynamic monitoring in various clinical conditions (USCOM, Ultrasonic cardiac output monitoring).
The present study aimed at comparing the accuracy of the USCOM device with that of the thermodilution technique in patients with septicemia.
We conducted a prospective observational study in a medical but noncardiological ICU of a university hospital.
Septic adult patients (median age 55 years, median SAPS-II-Score 43 points) on mechanical ventilation and catecholamine support were monitored with USCOM and PiCCO (n=70).
Seventy paired left-sided CO measurements (transaortic access = COUS-A) were obtained.
The mean COUS-A were 6.55 l/min (±2.19) versus COPiCCO 6.5 l/min (±2.18).
The correlation coefficient was r=0.89.
Comparison by Bland-Altman analysis revealed a bias of −0.36 l/min (±0.99 l/min) leading to a mean percentage error of 29%.
USCOM is a feasible and rapid method to evaluate CO in septic patients.
USCOM does reliably represent CO values as compared to the reference technique based on thermodilution (PiCCO).
It seems to be appropriate in situations where CO measurements are most pertinent to patient management.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Horster, Sophia& Stemmler, Hans-Joachim& Strecker, Nina& Brettner, Florian& Hausmann, Andreas& Cnossen, Jitske…[et al.]. 2011. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-459125
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Horster, Sophia…[et al.]. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-459125
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Horster, Sophia& Stemmler, Hans-Joachim& Strecker, Nina& Brettner, Florian& Hausmann, Andreas& Cnossen, Jitske…[et al.]. Cardiac Output Measurements in Septic Patients : Comparing the Accuracy of USCOM to PiCCO. Critical Care Research and Practice. 2011. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-459125
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-459125
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)