Comparative Evaluation of Two Types of Immediately Loaded Implants Using Biomechanical and Histomorphometric Tests : An Animal Case Study
المؤلفون المشاركون
Badrian, Hamid
Rismanchian, Mansour
Nekouie, Afsaneh
Khalighinejad, Navid
Movahedian, Bijan
Mohammad Razavi, Sayed
المصدر
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-07-18
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
Introduction.
In order to minimize the required time to regain esthetic and function, immediately loaded implants were suggested.
The aim of this study was to comparatively evaluate the Nisastan and XIve implants using biomechanical and histomorphometric tests.
Materials and Methods.
In this experimental study, 6 Nisastan one-piece immediately loaded screw type implant (OPILS) and 6 Xive implants with 3.4 mm diameter and 11 mm long were used.
The implants were immediately loaded with temporary coating.
After three months, the torque required to break bone-implant contact was measured and was recorded.
All implants were extracted with surrounding bone and histologically were evaluated.
The data were inputted into the SPSS 11.5 to run student T-test statistical analyses (α=0.05).
Results.
The success rates of both types of implants was 100%, and none of them failed due to mobility or bone loss.
The mean removal torque value (RTV) was 142.08 and 40 N/Cm for Xive and Nisastan implants, respectively, and their RTVs showed a significant difference between two mentioned implants (P=0.004).
None of the histomorphometric values showed significant differences between the two implants (P>0.05).
Discussion.
both systems have the capability to induce osseointegration under immediate loads but that Xive implants showed higher capability for bone contact.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Rismanchian, Mansour& Movahedian, Bijan& Khalighinejad, Navid& Badrian, Hamid& Mohammad Razavi, Sayed& Nekouie, Afsaneh. 2012. Comparative Evaluation of Two Types of Immediately Loaded Implants Using Biomechanical and Histomorphometric Tests : An Animal Case Study. ISRN Dentistry،Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-463964
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Rismanchian, Mansour…[et al.]. Comparative Evaluation of Two Types of Immediately Loaded Implants Using Biomechanical and Histomorphometric Tests : An Animal Case Study. ISRN Dentistry No. 2012 (2012), pp.1-6.
https://search.emarefa.net/detail/BIM-463964
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Rismanchian, Mansour& Movahedian, Bijan& Khalighinejad, Navid& Badrian, Hamid& Mohammad Razavi, Sayed& Nekouie, Afsaneh. Comparative Evaluation of Two Types of Immediately Loaded Implants Using Biomechanical and Histomorphometric Tests : An Animal Case Study. ISRN Dentistry. 2012. Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-463964
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-463964
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر