Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation

المؤلفون المشاركون

Schubert, Sarah A.
Pae, Walter E.
Soleimani, Behzad

المصدر

Case Reports in Transplantation

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-3، 3ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2012-09-26

دولة النشر

مصر

عدد الصفحات

3

التخصصات الرئيسية

الطب البشري

الملخص EN

We report a case of severe hemolysis and pulmonary valve insufficiency (PI) following right ventricular support using a paracorporeal pneumatic pump (Abiomed, Danvers, MA, USA).

We speculate that the high velocity jet of blood emanating from the outflow cannula caused turbulence above the pulmonary valve, leading to PI and hemolysis.

Despite the growing number of implanted ventricular assist devices, we could find no report in the literature describing pulmonary valve insufficiency secondary to right ventricular assist device (RVAD) placement.

Fortunately, in this case, right ventricular function recovered sufficiently after seven days of support, allowing explantation of the device and resolution of PI and hemolysis.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Schubert, Sarah A.& Soleimani, Behzad& Pae, Walter E.. 2012. Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation. Case Reports in Transplantation،Vol. 2012, no. 2012, pp.1-3.
https://search.emarefa.net/detail/BIM-467233

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Schubert, Sarah A.…[et al.]. Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation. Case Reports in Transplantation No. 2012 (2012), pp.1-3.
https://search.emarefa.net/detail/BIM-467233

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Schubert, Sarah A.& Soleimani, Behzad& Pae, Walter E.. Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation. Case Reports in Transplantation. 2012. Vol. 2012, no. 2012, pp.1-3.
https://search.emarefa.net/detail/BIM-467233

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-467233