الملخص EN

A fractured central venous catheter (CVC) with embolization of the distal fragment may lead to life-threatening complications.

We had inserted a right subclavian CVC in a 68-year-old female which upon a follow-up chest X-ray appeared to have been sheared.

A guidewire was inserted through the CVC until the J-tip was just beyond the tip of the CVC which were then withdrawn as a single assembly.

We suspected that the tip of the guidewire might have been entrapped in the opening of the middle port, which upon withdrawal of the guidewire could have led to CVC folding upon itself and shearing.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Hegde, Harihar V.& Yaliwal, Vijay G.& Joshi, Shyamsundar K.& Rao, P. Raghavendra. 2011. The Sheared Central Venous Catheter?. Case Reports in Anesthesiology،Vol. 2011, no. 2011, pp.1-4.
https://search.emarefa.net/detail/BIM-467447

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Hegde, Harihar V.…[et al.]. The Sheared Central Venous Catheter?. Case Reports in Anesthesiology No. 2011 (2011), pp.1-4.
https://search.emarefa.net/detail/BIM-467447

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Hegde, Harihar V.& Yaliwal, Vijay G.& Joshi, Shyamsundar K.& Rao, P. Raghavendra. The Sheared Central Venous Catheter?. Case Reports in Anesthesiology. 2011. Vol. 2011, no. 2011, pp.1-4.
https://search.emarefa.net/detail/BIM-467447

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-467447