Fibre Optic Readout of Microcantilever Arrays for Fast Microorganism Growth Detection

المؤلفون المشاركون

Jensen, Jason
Grange, Wilfried
Ramseyer, J. P.
Maloney, N.
Nugaeva, N.
Hegner, Martin
Lukacs, Gyongyi

المصدر

Journal of Sensors

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-09-13

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

هندسة مدنية

الملخص EN

We present a fibre-optic-based device for the automated readout of microcantilever arrays for fast microorganism growth detection.

We determined the ability of our device to track shifts in resonance frequency due to an increase in mass on the cantilever surface or changes in mechanical stiffness.

The resonance frequency response of 7 μm thick agarose-functionalised cantilevers was tracked as humidity levels were varied revealing a mass responsivity of ~51±1 pg/Hz.

The resonance response of microcantilevers coated with Aspergillus niger (A.

niger) spores was monitored for >48 h revealing a growth detection time of >4 h.

The growth of mycelium along the cantilevers surface is seen to result in an increase in resonance frequency due to the reinforcement of the cantilever structure.

The use of our fibre optic detection technique allows data to be recorded continuously and faster than previously reported.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Maloney, N.& Lukacs, Gyongyi& Nugaeva, N.& Grange, Wilfried& Ramseyer, J. P.& Jensen, Jason…[et al.]. 2011. Fibre Optic Readout of Microcantilever Arrays for Fast Microorganism Growth Detection. Journal of Sensors،Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-469458

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Maloney, N.…[et al.]. Fibre Optic Readout of Microcantilever Arrays for Fast Microorganism Growth Detection. Journal of Sensors No. 2012 (2012), pp.1-6.
https://search.emarefa.net/detail/BIM-469458

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Maloney, N.& Lukacs, Gyongyi& Nugaeva, N.& Grange, Wilfried& Ramseyer, J. P.& Jensen, Jason…[et al.]. Fibre Optic Readout of Microcantilever Arrays for Fast Microorganism Growth Detection. Journal of Sensors. 2011. Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-469458

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-469458