Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus

المؤلفون المشاركون

Mulcahy, Daniel M.
Stoskopf, Michael K.
Esler, Daniel

المصدر

Veterinary Medicine International

العدد

المجلد 2010، العدد 2010 (31 ديسمبر/كانون الأول 2010)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2010-02-22

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

علم الحيوان

الملخص EN

A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska.

We compared serum chemistry results obtained on-site with results from a traditional laboratory.

Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data.

The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site.

Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L).

Serum potassium values were too low for quantitation by the traditional laboratory.

Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Stoskopf, Michael K.& Mulcahy, Daniel M.& Esler, Daniel. 2010. Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus. Veterinary Medicine International،Vol. 2010, no. 2010, pp.1-5.
https://search.emarefa.net/detail/BIM-470654

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Stoskopf, Michael K.…[et al.]. Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus. Veterinary Medicine International No. 2010 (2010), pp.1-5.
https://search.emarefa.net/detail/BIM-470654

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Stoskopf, Michael K.& Mulcahy, Daniel M.& Esler, Daniel. Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus. Veterinary Medicine International. 2010. Vol. 2010, no. 2010, pp.1-5.
https://search.emarefa.net/detail/BIM-470654

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-470654