Respiratory-Gated MRgHIFU in Upper Abdomen Using an MR-Compatible In-Bore Digital Camera
المؤلفون المشاركون
Petrusca, Lorena
Becker, Christoph D.
Salomir, Rares
Auboiroux, Vincent
Viallon, Magalie
Montet, Xavier
Muller, Arnaud
Breguet, Romain
Terraz, Sylvain
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-01-29
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
Objective.
To demonstrate the technical feasibility and the potential interest of using a digital optical camera inside the MR magnet bore for monitoring the breathing cycle and subsequently gating the PRFS MR thermometry, MR-ARFI measurement, and MRgHIFU sonication in the upper abdomen.
Materials and Methods.
A digital camera was reengineered to remove its magnetic parts and was further equipped with a 7 m long USB cable.
The system was electromagnetically shielded and operated inside the bore of a closed 3T clinical scanner.
Suitable triggers were generated based on real-time motion analysis of the images produced by the camera (resolution 640×480 pixels, 30 fps).
Respiratory-gated MR-ARFI prepared MRgHIFU ablation was performed in the kidney and liver of two sheep in vivo, under general anaesthesia and ventilator-driven forced breathing.
Results.
The optical device demonstrated very good MR compatibility.
The current setup permitted the acquisition of motion artefact-free and high resolution MR 2D ARFI and multiplanar interleaved PRFS thermometry (average SNR 30 in liver and 56 in kidney).
Microscopic histology indicated precise focal lesions with sharply delineated margins following the respiratory-gated HIFU sonications.
Conclusion.
The proof-of-concept for respiratory motion management in MRgHIFU using an in-bore digital camera has been validated in vivo.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Auboiroux, Vincent& Petrusca, Lorena& Viallon, Magalie& Muller, Arnaud& Terraz, Sylvain& Breguet, Romain…[et al.]. 2014. Respiratory-Gated MRgHIFU in Upper Abdomen Using an MR-Compatible In-Bore Digital Camera. BioMed Research International،Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-470956
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Auboiroux, Vincent…[et al.]. Respiratory-Gated MRgHIFU in Upper Abdomen Using an MR-Compatible In-Bore Digital Camera. BioMed Research International No. 2014 (2014), pp.1-9.
https://search.emarefa.net/detail/BIM-470956
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Auboiroux, Vincent& Petrusca, Lorena& Viallon, Magalie& Muller, Arnaud& Terraz, Sylvain& Breguet, Romain…[et al.]. Respiratory-Gated MRgHIFU in Upper Abdomen Using an MR-Compatible In-Bore Digital Camera. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-470956
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-470956
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"