الملخص EN
Hip fractures often result in serious health implications, particularly in the geriatric population, and have been related to long-term morbidity and death.
In most cases, these fractures are caused by impact loads in the area of the greater trochanter, which are produced in a fall.
This work is aimed at developing hip protectors using composite materials and evaluating their effectiveness in preventing hip fractures under high impact energy (120 J).
The hip protectors were developed with an inner layer of energy absorbing soft material and an outer rigid shell of fiberglass-reinforced polymer composite.
According to the experimental results, all tested configurations proved to be effective at reducing the impact load to below the average fracture threshold of proximal femur.
Furthermore, an addition of Ethylene Vinyl Acetate (EVA) to the impacted area of the composite shell proved to be beneficial to increase impact strength of the hip protectors.
Thus, composite hip protectors proved to be a viable alternative for a mechanically efficient and cost-effective solution to prevent hip fractures.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Melo, Jose Daniel Diniz& Barbosa, Ayrles S. Gonçalves& Oliveira Guerra, Ricardo. 2010. Mechanical Evaluation of Polymer Composite Hip Protectors. International Journal of Biomaterials،Vol. 2010, no. 2010, pp.1-6.
https://search.emarefa.net/detail/BIM-471760
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Melo, Jose Daniel Diniz…[et al.]. Mechanical Evaluation of Polymer Composite Hip Protectors. International Journal of Biomaterials No. 2010 (2010), pp.1-6.
https://search.emarefa.net/detail/BIM-471760
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Melo, Jose Daniel Diniz& Barbosa, Ayrles S. Gonçalves& Oliveira Guerra, Ricardo. Mechanical Evaluation of Polymer Composite Hip Protectors. International Journal of Biomaterials. 2010. Vol. 2010, no. 2010, pp.1-6.
https://search.emarefa.net/detail/BIM-471760
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-471760
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر