Despite the use of advanced radiological investigations, some liver lesions cannot be definitely diagnosed without a biopsy and histological examination.
Laparoscopic Tru-Cut biopsy of the liver lesion is the preferred approach to achieve a good sample for histology.
The mechanism of a Tru-Cut biopsy needle needs the use of both hands to load and fire the needle.
This restricts the ability of the surgeon to direct the needle into the lesion utilising the laparoscopic ultrasound probe.
We report a technique of laparoscopic liver biopsy using a disposable core biopsy instrument (BARD (R) disposable core biopsy needle) that can be used single-handedly.
The needle can be positioned with laparoscopic graspers in order to reach posterior and superior lesions.
This technique can easily be used in conjunction with laparoscopic ultrasound.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Trochsler, Markus& Ralph, Q.& Bridgewater, F.& Kanhere, Harsh A.& Maddern, Guy J.. 2013. Technical Note : Facilitating Laparoscopic Liver Biopsy by the Use of a Single-Handed Disposable Core Biopsy Needle. HPB Surgery،Vol. 2013, no. 2013, pp.1-4.
https://search.emarefa.net/detail/BIM-473481
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Trochsler, Markus…[et al.]. Technical Note : Facilitating Laparoscopic Liver Biopsy by the Use of a Single-Handed Disposable Core Biopsy Needle. HPB Surgery No. 2013 (2013), pp.1-4.
https://search.emarefa.net/detail/BIM-473481
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Trochsler, Markus& Ralph, Q.& Bridgewater, F.& Kanhere, Harsh A.& Maddern, Guy J.. Technical Note : Facilitating Laparoscopic Liver Biopsy by the Use of a Single-Handed Disposable Core Biopsy Needle. HPB Surgery. 2013. Vol. 2013, no. 2013, pp.1-4.
https://search.emarefa.net/detail/BIM-473481
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-473481
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)