Isolated Atrial Lead Conduction Delay following Right Atrial Radiofrequency Maze Procedure

المؤلفون المشاركون

Dearani, Joseph A.
Shen, Win-Kuang
Eleid, Mackram F.

المصدر

ISRN Cardiology

العدد

المجلد 2011، العدد 2011 (31 ديسمبر/كانون الأول 2011)، ص ص. 1-3، 3ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-06-01

دولة النشر

مصر

عدد الصفحات

3

التخصصات الرئيسية

الأمراض

الملخص EN

A 60-year-old man with a dual-chamber implantable cardioverter defibrillator and severe dyspnea on exertion due to apical hypertrophic cardiomyopathy underwent a septal myectomy and radiofrequency maze procedure.

Following the procedure a persistent delay in atrial sensing was observed and was most likely a result of iatrogenic conduction delay from right atrial ablation lines.

These observations suggest that atrial conduction properties can be altered during the surgical maze procedure and should be considered in the differential diagnosis of sensing or pacing malfunction.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Eleid, Mackram F.& Dearani, Joseph A.& Shen, Win-Kuang. 2011. Isolated Atrial Lead Conduction Delay following Right Atrial Radiofrequency Maze Procedure. ISRN Cardiology،Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-474559

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Eleid, Mackram F.…[et al.]. Isolated Atrial Lead Conduction Delay following Right Atrial Radiofrequency Maze Procedure. ISRN Cardiology No. 2011 (2011), pp.1-3.
https://search.emarefa.net/detail/BIM-474559

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Eleid, Mackram F.& Dearani, Joseph A.& Shen, Win-Kuang. Isolated Atrial Lead Conduction Delay following Right Atrial Radiofrequency Maze Procedure. ISRN Cardiology. 2011. Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-474559

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-474559