Correcting Inaccurately Recorded Data due to Faulty Calibration of a Capacitance Water Content Probe

المؤلفون المشاركون

ElNesr, Mohammad N.
El-Farrah, Mohammad A.
Alazba, A. A.

المصدر

Applied and Environmental Soil Science

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-10، 10ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2013-08-05

دولة النشر

مصر

عدد الصفحات

10

التخصصات الرئيسية

علم الأرض والمياه والبيئة

الملخص EN

Measuring soil water content by capacitance probes requires rigorous calibration to achieve acceptable accuracy.

Some of the capacitance probes' users might take several readings using the default device calibrations or other prestored calibrations by mistake.

This can lead to logging of faulty readings for periods of up to months or years.

This study aimed to (1) study the importance of probe calibration and the level of error that results from using flawed calibrations and (2) to develop a mathematical method to correct the faulty recorded data.

This research involved studying eleven scenarios of faulty calibrations including errors in the air/water calibration and in the in-soil calibration.

A mathematical method was developed to correct the faulty recorded data and comparisons were made for the data after and before correction.

Results indicated that using the manufacturer's default calibration within the software resulted in substantial error values especially for heavy textured soils.

It is recommended that users and especially researchers should perform rigorous in-soil calibration wherever the probe is installed, and they should repeat the calibrations whenever the soil structure changed.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

ElNesr, Mohammad N.& Alazba, A. A.& El-Farrah, Mohammad A.. 2013. Correcting Inaccurately Recorded Data due to Faulty Calibration of a Capacitance Water Content Probe. Applied and Environmental Soil Science،Vol. 2013, no. 2013, pp.1-10.
https://search.emarefa.net/detail/BIM-479115

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

ElNesr, Mohammad N.…[et al.]. Correcting Inaccurately Recorded Data due to Faulty Calibration of a Capacitance Water Content Probe. Applied and Environmental Soil Science No. 2013 (2013), pp.1-10.
https://search.emarefa.net/detail/BIM-479115

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

ElNesr, Mohammad N.& Alazba, A. A.& El-Farrah, Mohammad A.. Correcting Inaccurately Recorded Data due to Faulty Calibration of a Capacitance Water Content Probe. Applied and Environmental Soil Science. 2013. Vol. 2013, no. 2013, pp.1-10.
https://search.emarefa.net/detail/BIM-479115

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-479115