Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens
المؤلفون المشاركون
De Seta, Daniele
Sterkers, Olivier
Miroir, Mathieu
Ferrary, Evelyne
Kazmitcheff, Guillaume
Nguyen, Yann
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-07-14
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
Introduction.
In order to achieve a minimal trauma to the inner ear structures during array insertion, it would be suitable to control insertion forces.
The aim of this work was to compare the insertion forces of an array insertion into anatomical specimens with three different insertion techniques: with forceps, with a commercial tool, and with a motorized tool.
Materials and Methods.
Temporal bones have been mounted on a 6-axis force sensor to record insertion forces.
Each temporal bone has been inserted, with a lateral wall electrode array, in random order, with each of the 3 techniques.
Results.
Forceps manual and commercial tool insertions generated multiple jerks during whole length insertion related to fits and starts.
On the contrary, insertion force with the motorized tool only rose at the end of the insertion.
Overall force momentum was 1.16 ± 0.505 N (mean ± SD, n=10), 1.337 ± 0.408 N (n=8), and 1.573 ± 0.764 N (n=8) for manual insertion with forceps and commercial and motorized tools, respectively.
Conclusion.
Considering force momentum, no difference between the three techniques was observed.
Nevertheless, a more predictable force profile could be observed with the motorized tool with a smoother rise of insertion forces.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Nguyen, Yann& Kazmitcheff, Guillaume& De Seta, Daniele& Miroir, Mathieu& Ferrary, Evelyne& Sterkers, Olivier. 2014. Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens. BioMed Research International،Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-479272
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Nguyen, Yann…[et al.]. Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens. BioMed Research International No. 2014 (2014), pp.1-9.
https://search.emarefa.net/detail/BIM-479272
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Nguyen, Yann& Kazmitcheff, Guillaume& De Seta, Daniele& Miroir, Mathieu& Ferrary, Evelyne& Sterkers, Olivier. Definition of Metrics to Evaluate Cochlear Array Insertion Forces Performed with Forceps, Insertion Tool, or Motorized Tool in Temporal Bone Specimens. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-479272
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-479272
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)