An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation
المؤلفون المشاركون
Hu, Xiaodong
Hu, Xiaotang
Lv, Hanyu
Zou, Jing
المصدر
International Journal of Biomedical Imaging
العدد
المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-8، 8ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2013-12-25
دولة النشر
مصر
عدد الصفحات
8
التخصصات الرئيسية
الملخص EN
The effects of calibration phantoms on the correction results of the empirical artifacts correction method (ECCU) for the case of tube modulation were investigated.
To improve the validity of the ECCU method, the effect of the geometry parameter of a typical single-material calibration phantom (water calibration phantom) on the ECCU algorithm was investigated.
Dual-material calibration phantoms (such as water-bone calibration phantom), geometry arrangement, and the area-ratio of dual-material calibration phantoms were also studied.
Preliminary results implied that, to assure the effectiveness of the ECCU algorithm, the polychromatic projections of calibration phantoms must cover the polychromatic projection data of the scanning object.
However, the projection range of a water calibration phantom is limited by the scan field of view (SFOV), thus leading to methodological limitations.
A dual-material phantom of a proper size and material can overcome the limitations of a single-material phantom and achieve good correction effects.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zou, Jing& Hu, Xiaodong& Lv, Hanyu& Hu, Xiaotang. 2013. An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation. International Journal of Biomedical Imaging،Vol. 2013, no. 2013, pp.1-8.
https://search.emarefa.net/detail/BIM-481063
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zou, Jing…[et al.]. An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation. International Journal of Biomedical Imaging No. 2013 (2013), pp.1-8.
https://search.emarefa.net/detail/BIM-481063
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zou, Jing& Hu, Xiaodong& Lv, Hanyu& Hu, Xiaotang. An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation. International Journal of Biomedical Imaging. 2013. Vol. 2013, no. 2013, pp.1-8.
https://search.emarefa.net/detail/BIM-481063
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-481063
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)