Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM) : An Ovine Cardiac Output Validation
المؤلفون المشاركون
West, Malcolm J.
Wan, Li
Hood, Sally G.
Jacobson, Beverley M.
Phillips, Robert A.
May, Clive N.
المصدر
Critical Care Research and Practice
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-05-09
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
Background.
The pulmonary artery catheter (PAC) is an accepted clinical method of measuring cardiac output (CO) despite no prior validation.
The ultrasonic cardiac output monitor (USCOM) is a noninvasive alternative to PAC using Doppler ultrasound (CW).
We compared PAC and USCOM CO measurements against a gold standard, the aortic flow probe (FP), in sheep at varying outputs.
Methods.
Ten conscious sheep, with implanted FPs, had measurements of CO by FP, USCOM, and PAC, at rest and during intervention with inotropes and vasopressors.
Results.
CO measurements by FP, PAC, and USCOM were 4.0±1.2 L/min, 4.8±1.5 L/min, and 4.0±1.4 L/min, respectively, (n=280, range 1.9 L/min to 11.7 L/min).
Percentage bias and precision between FP and PAC, and FP and USCOM was −17 and 47%, and 1 and 36%, respectively.
PAC under-measured Dobutamine-induced CO changes by 20% (relative 66%) compared with FP, while USCOM measures varied from FP by 3% (relative 10%).
PAC reliably detected −30% but not +40% CO changes, as measured by receiver operating characteristic area under the curve (AUC), while USCOM reliably detected ±5% changes in CO (AUC>0.70).
Conclusions.
PAC demonstrated poor accuracy and sensitivity as a measure of CO.
USCOM provided equivalent measurements to FP across a sixfold range of outputs, reliably detecting ±5% changes.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Phillips, Robert A.& Hood, Sally G.& Jacobson, Beverley M.& West, Malcolm J.& Wan, Li& May, Clive N.. 2012. Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM) : An Ovine Cardiac Output Validation. Critical Care Research and Practice،Vol. 2012, no. 2012, pp.1-9.
https://search.emarefa.net/detail/BIM-485861
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Phillips, Robert A.…[et al.]. Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM) : An Ovine Cardiac Output Validation. Critical Care Research and Practice No. 2012 (2012), pp.1-9.
https://search.emarefa.net/detail/BIM-485861
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Phillips, Robert A.& Hood, Sally G.& Jacobson, Beverley M.& West, Malcolm J.& Wan, Li& May, Clive N.. Pulmonary Artery Catheter (PAC) Accuracy and Efficacy Compared with Flow Probe and Transcutaneous Doppler (USCOM) : An Ovine Cardiac Output Validation. Critical Care Research and Practice. 2012. Vol. 2012, no. 2012, pp.1-9.
https://search.emarefa.net/detail/BIM-485861
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-485861
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)