Clinical Outcomes of Tibial Components with Modular Stems Used in Primary TKA

المؤلفون المشاركون

Harman, Melinda K.
Broome, Brandon
Osuji, Obi
Durig, Nicole
Pace, Thomas B.

المصدر

Advances in Orthopedics

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-02-05

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Due to the known potential for fretting and corrosion at modular junctions in orthopaedic implants, this retrospective study evaluated radiographic and clinical outcomes of 85 primary TKA patients implanted with modular stemmed tibial components and followed up for an average of 82 months.

There was low incidence of tibial radiolucent lines, excellent functional outcomes, and no complications associated with stem modularity.

The findings were comparable to the historical control study involving 107 TKA with a nonmodular tibial stem design.

When using surface cemented tibial components combined with a constrained polyethylene bearing, modular stems appear to be a viable option for primary TKA when adequate fixation and rotational stability are maintained.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Durig, Nicole& Pace, Thomas B.& Broome, Brandon& Osuji, Obi& Harman, Melinda K.. 2014. Clinical Outcomes of Tibial Components with Modular Stems Used in Primary TKA. Advances in Orthopedics،Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-488297

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Durig, Nicole…[et al.]. Clinical Outcomes of Tibial Components with Modular Stems Used in Primary TKA. Advances in Orthopedics No. 2014 (2014), pp.1-6.
https://search.emarefa.net/detail/BIM-488297

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Durig, Nicole& Pace, Thomas B.& Broome, Brandon& Osuji, Obi& Harman, Melinda K.. Clinical Outcomes of Tibial Components with Modular Stems Used in Primary TKA. Advances in Orthopedics. 2014. Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-488297

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل