Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis

المؤلفون المشاركون

Vaidya, Rahul
Roth, Matthew
Onwudiwe, Ndidi
Sethi, Anil

المصدر

Advances in Orthopedics

العدد

المجلد 2013، العدد 2013 (31 ديسمبر/كانون الأول 2013)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2013-12-03

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

Comparison of monoaxial and polyaxial screws with the use of subcutaneous anterior pelvic fixation.

Methods.

Four different groups each having 5 constructs were tested in distraction within the elastic range.

Once that was completed, 3 components were tested in torsion within the elastic range, 2 to torsional failure and 3 in distraction until failure.

Results.

The pedicle screw systems showed higher stiffness (4.008 ± 0.113 Nmm monoaxial, 3.638 ± 0.108 Nmm Click-x; 3.634 ± 0.147 Nmm Pangea) than the exfix system (2.882 ± 0.054 Nmm) in distraction.

In failure testing, monoaxial pedicle screw system was stronger (360 N) than exfixes (160 N) and polyaxial devices which failed if distracted greater than 4 cm (157 N Click-x or 138 N Pangea).

The exfix had higher peak torque and torsional stiffness than all pedicle systems.

In torsion, the yield strengths were the same for all constructs.

Conclusion.

The infix device constructed with polyaxial or monoaxial pedicle screws is stiffer than the 2 pin external fixator in distraction testing.

In extreme cases, the use of reinforcement or monoaxial systems which do not fail even at 360 N is a better option.

In torsional testing, the 2 pin external fixator is stiffer than the pedicle screw systems.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Vaidya, Rahul& Onwudiwe, Ndidi& Roth, Matthew& Sethi, Anil. 2013. Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis. Advances in Orthopedics،Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-490225

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Vaidya, Rahul…[et al.]. Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis. Advances in Orthopedics No. 2013 (2013), pp.1-6.
https://search.emarefa.net/detail/BIM-490225

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Vaidya, Rahul& Onwudiwe, Ndidi& Roth, Matthew& Sethi, Anil. Monoaxial Pedicle Screws Are Superior to Polyaxial Pedicle Screws and the Two Pin External Fixator for Subcutaneous Anterior Pelvic Fixation in a Biomechanical Analysis. Advances in Orthopedics. 2013. Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-490225

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-490225