Evaluating the Reproducibility of Motion Analysis Scanning of the Spine during Walking
المؤلفون المشاركون
Gipsman, Aaron
Daneshvar, Michael
Knott, Patrick
Rauschert, Lisa
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-9، 9ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-07-17
دولة النشر
مصر
عدد الصفحات
9
التخصصات الرئيسية
الملخص EN
The Formetric 4D dynamic system (Diers International GmbH, Schlangenbad, Germany) is a rasterstereography based imaging system designed to evaluate spinal deformity, providing radiation-free imaging of the position, rotation, and shape of the spine during the gait cycle.
Purpose.
This study was designed to evaluate whether repeated measurements with the Formetric 4D dynamic system would be reproducible with a standard deviation of less than +/− 3 degrees.
This study looked at real-time segmental motion, measuring kyphosis, lordosis, trunk length, pelvic, and T4 and L1 vertebral body rotation.
Methods.
Twenty healthy volunteers each underwent 3 consecutive scans.
Measurements for kyphosis, lordosis, trunk length, and rotations of T4, L1, and the pelvis were recorded for each trial.
Results.
The average standard deviations of same-day repeat measurements were within +/− 3 degrees with a range of 0.51 degrees to 2.3 degrees.
Conclusions.
The surface topography system calculated reproducible measurements with error ranges comparable to the current gold standard in dynamic spinal motion analysis.
Therefore, this technique should be considered of high clinical value for reliably evaluating segmental motion and spinal curvatures and should further be evaluated in the setting of adolescent idiopathic scoliosis.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Gipsman, Aaron& Rauschert, Lisa& Daneshvar, Michael& Knott, Patrick. 2014. Evaluating the Reproducibility of Motion Analysis Scanning of the Spine during Walking. Advances in Medicine،Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-493378
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Gipsman, Aaron…[et al.]. Evaluating the Reproducibility of Motion Analysis Scanning of the Spine during Walking. Advances in Medicine No. 2014 (2014), pp.1-9.
https://search.emarefa.net/detail/BIM-493378
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Gipsman, Aaron& Rauschert, Lisa& Daneshvar, Michael& Knott, Patrick. Evaluating the Reproducibility of Motion Analysis Scanning of the Spine during Walking. Advances in Medicine. 2014. Vol. 2014, no. 2014, pp.1-9.
https://search.emarefa.net/detail/BIM-493378
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-493378
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر