Radiation Dose Reduction during Radial Cardiac Catheterization : Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape
المؤلفون المشاركون
Sweis, Ranya
Nadelson, Jeffrey
Shroff, Adhir R.
Ferrera, Dean
Vidovich, Mladen I.
Ertel, Andrew
المصدر
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2012-09-04
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
Objectives.
Radiation scatter protection shield drapes have been designed with the goal of decreasing radiation dose to the operators during transfemoral catheterization.
We sought to investigate the impact on operator radiation exposure of various shielding drapes specifically designed for the radial approach.
Background.
Radial access for cardiac catheterization has increased due to improved patient comfort and decreased bleeding complications.
There are concerns for increased radiation exposure to patients and operators.
Methods.
Radiation doses to a simulated operator were measured with a RadCal Dosimeter in the cardiac catheterization laboratory.
The mock patient was a 97.5 kg fission product phantom.
Three lead-free drape designs were studied.
The drapes were placed just proximal to the right wrist and extended medially to phantom’s trunk.
Simulated diagnostic coronary angiography included 6 minutes of fluoroscopy time and 32 seconds of cineangiography time at 4 standard angulated views (8 s each), both 15 frames/s.
ANOVA with Bonferroni correction was used for statistical analysis.
Results.
All drape designs led to substantial reductions in operator radiation exposure compared to control (P<0.0001).
The greatest decrease in radiation exposure (72%) was with the L-shaped design.
Conclusions.
Dedicated radial shielding drapes decrease radiation exposure to the operator by up to 72% during simulated cardiac catheterization.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Ertel, Andrew& Nadelson, Jeffrey& Shroff, Adhir R.& Sweis, Ranya& Ferrera, Dean& Vidovich, Mladen I.. 2012. Radiation Dose Reduction during Radial Cardiac Catheterization : Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape. ISRN Cardiology،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-497312
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Ertel, Andrew…[et al.]. Radiation Dose Reduction during Radial Cardiac Catheterization : Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape. ISRN Cardiology No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-497312
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Ertel, Andrew& Nadelson, Jeffrey& Shroff, Adhir R.& Sweis, Ranya& Ferrera, Dean& Vidovich, Mladen I.. Radiation Dose Reduction during Radial Cardiac Catheterization : Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape. ISRN Cardiology. 2012. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-497312
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-497312
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر