Human Heart Pulse Wave Responses Measured Simultaneously at Several Sensor Placements by Two MR-Compatible Fibre Optic Methods

المؤلفون المشاركون

Korhonen, Vesa
Sorvoja, Hannu
Tervonen, Osmo
Hiltunen, Tuija
Vihriälä, Erkki
Kiviniemi, Vesa
Myllylä, Teemu

المصدر

Journal of Sensors

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-8، 8ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2012-12-17

دولة النشر

مصر

عدد الصفحات

8

التخصصات الرئيسية

هندسة مدنية

الملخص EN

This paper presents experimental measurements conducted using two noninvasive fibre optic methods for detecting heart pulse waves in the human body.

Both methods can be used in conjunction with magnetic resonance imaging (MRI).

For comparison, the paper also performs an MRI-compatible electrocardiogram (ECG) measurement.

By the simultaneous use of different measurement methods, the propagation of pressure waves generated by each heart pulse can be sensed extensively in different areas of the human body and at different depths, for example, on the chest and forehead and at the fingertip.

An accurate determination of a pulse wave allows calculating the pulse transit time (PTT) of a particular heart pulse in different parts of the human body.

This result can then be used to estimate the pulse wave velocity of blood flow in different places.

Both measurement methods are realized using magnetic resonance-compatible fibres, which makes the methods applicable to the MRI environment.

One of the developed sensors is an extraordinary accelerometer sensor, while the other one is a more common sensor based on photoplethysmography.

All measurements, involving several test patients, were performed both inside and outside an MRI room.

Measurements inside the MRI room were conducted using a 3-Tesla strength closed MRI scanner in the Department of Diagnostic Radiology at the Oulu University Hospital.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Myllylä, Teemu& Korhonen, Vesa& Vihriälä, Erkki& Sorvoja, Hannu& Hiltunen, Tuija& Tervonen, Osmo…[et al.]. 2012. Human Heart Pulse Wave Responses Measured Simultaneously at Several Sensor Placements by Two MR-Compatible Fibre Optic Methods. Journal of Sensors،Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-497358

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Myllylä, Teemu…[et al.]. Human Heart Pulse Wave Responses Measured Simultaneously at Several Sensor Placements by Two MR-Compatible Fibre Optic Methods. Journal of Sensors No. 2012 (2012), pp.1-8.
https://search.emarefa.net/detail/BIM-497358

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Myllylä, Teemu& Korhonen, Vesa& Vihriälä, Erkki& Sorvoja, Hannu& Hiltunen, Tuija& Tervonen, Osmo…[et al.]. Human Heart Pulse Wave Responses Measured Simultaneously at Several Sensor Placements by Two MR-Compatible Fibre Optic Methods. Journal of Sensors. 2012. Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-497358

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-497358