Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes

المؤلفون المشاركون

Pinilla, Isabel
Lopez de la Fuente, Carmen
Fuentes-Broto, Lorena
Sanchez-Cano, Ana
Segura, Francisco

المصدر

BioMed Research International

العدد

المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-6، 6ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2014-04-24

دولة النشر

مصر

عدد الصفحات

6

التخصصات الرئيسية

الطب البشري

الملخص EN

Purpose.

To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination.

Methods.

Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited.

Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment.

Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV).

Results.

Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60).

Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities.

In these last three parameters, repeatability limits were excessively high compared to the mean.

Conclusions.

GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms.

Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited.

In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Lopez de la Fuente, Carmen& Sanchez-Cano, Ana& Segura, Francisco& Fuentes-Broto, Lorena& Pinilla, Isabel. 2014. Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes. BioMed Research International،Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-499760

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Lopez de la Fuente, Carmen…[et al.]. Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes. BioMed Research International No. 2014 (2014), pp.1-6.
https://search.emarefa.net/detail/BIM-499760

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Lopez de la Fuente, Carmen& Sanchez-Cano, Ana& Segura, Francisco& Fuentes-Broto, Lorena& Pinilla, Isabel. Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-499760

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-499760