![](/images/graphics-bg.png)
Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry
المؤلفون المشاركون
Haenle, Maximilian
Fritsche, Andreas
Finke, Birgit
Bader, Rainer
Zietz, Carmen
Stranak, Vitezslav
Mittelmeier, Wolfram
Hippler, Rainer
المصدر
Bioinorganic Chemistry and Applications
العدد
المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2011-11-24
دولة النشر
مصر
عدد الصفحات
5
التخصصات الرئيسية
الملخص EN
New developments of antimicrobial implant surfaces doped with copper (Cu) ions may minimize the risk of implant-associated infections.
However, experimental evaluation of the Cu release is influenced by various test parameters.
The aim of our study was to evaluate the Cu release characteristics in vitro according to the storage fluid and surface roughness.
Plasma immersion ion implantation of Cu (Cu-PIII) and pulsed magnetron sputtering process of a titanium copper film (Ti-Cu) were applied to titanium alloy (Ti6Al4V) samples with different surface finishing of the implant material (polished, hydroxyapatite and corundum blasted).
The samples were submersed into either double-distilled water, human serum, or cell culture medium.
Subsequently, the Cu concentration in the supernatant was measured using atomic absorption spectrometry.
The test fluid as well as the surface roughness can alter the Cu release significantly, whereby the highest Cu release was determined for samples with corundum-blasted surfaces stored in cell medium.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Zietz, Carmen& Fritsche, Andreas& Finke, Birgit& Stranak, Vitezslav& Haenle, Maximilian& Hippler, Rainer…[et al.]. 2011. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-503207
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Zietz, Carmen…[et al.]. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-503207
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Zietz, Carmen& Fritsche, Andreas& Finke, Birgit& Stranak, Vitezslav& Haenle, Maximilian& Hippler, Rainer…[et al.]. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications. 2011. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-503207
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-503207
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
![](/images/ebook-kashef.png)
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر
![](/images/kashef-image.png)