Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry

المؤلفون المشاركون

Haenle, Maximilian
Fritsche, Andreas
Finke, Birgit
Bader, Rainer
Zietz, Carmen
Stranak, Vitezslav
Mittelmeier, Wolfram
Hippler, Rainer

المصدر

Bioinorganic Chemistry and Applications

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2011-11-24

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الأحياء

الملخص EN

New developments of antimicrobial implant surfaces doped with copper (Cu) ions may minimize the risk of implant-associated infections.

However, experimental evaluation of the Cu release is influenced by various test parameters.

The aim of our study was to evaluate the Cu release characteristics in vitro according to the storage fluid and surface roughness.

Plasma immersion ion implantation of Cu (Cu-PIII) and pulsed magnetron sputtering process of a titanium copper film (Ti-Cu) were applied to titanium alloy (Ti6Al4V) samples with different surface finishing of the implant material (polished, hydroxyapatite and corundum blasted).

The samples were submersed into either double-distilled water, human serum, or cell culture medium.

Subsequently, the Cu concentration in the supernatant was measured using atomic absorption spectrometry.

The test fluid as well as the surface roughness can alter the Cu release significantly, whereby the highest Cu release was determined for samples with corundum-blasted surfaces stored in cell medium.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Zietz, Carmen& Fritsche, Andreas& Finke, Birgit& Stranak, Vitezslav& Haenle, Maximilian& Hippler, Rainer…[et al.]. 2011. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-503207

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Zietz, Carmen…[et al.]. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-503207

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Zietz, Carmen& Fritsche, Andreas& Finke, Birgit& Stranak, Vitezslav& Haenle, Maximilian& Hippler, Rainer…[et al.]. Analysis of the Release Characteristics of Cu-Treated Antimicrobial Implant Surfaces Using Atomic Absorption Spectrometry. Bioinorganic Chemistry and Applications. 2011. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-503207

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references