Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry

المؤلفون المشاركون

Kaluzny, Jozef
Burduk, Mateusz
Kaluzny, Bartlomiej J.
Soldanska, Beata
Piotrowiak, Ilona

المصدر

ISRN Ophthalmology

العدد

المجلد 2012، العدد 2012 (31 ديسمبر/كانون الأول 2012)، ص ص. 1-5، 5ص.

الناشر

Hindawi Publishing Corporation

تاريخ النشر

2012-09-27

دولة النشر

مصر

عدد الصفحات

5

التخصصات الرئيسية

الطب البشري

الملخص EN

Background and Objective.

Evaluation of agreement, repeatability, and reproducibility of central and minimal corneal thickness (CCT and MCT) measurements obtained by SOCT, the Scheimpflug system, and ultrasound pachymetry.

Materials and Methods.

28 eyes of healthy patients were enrolled.

Pachymetry measurements were performed with SOCT, the Scheimpflug system, and ultrasound instrument.

Each measurement was taken by 3 operators on 3 devices providing a total of 2100 measurements.

Results.

The mean CCT for SOCT, Scheimpflug system, and ultrasound instrument was 537.92, 545.94, and 555.74 μm, respectively, (P<.001).

The respective mean coefficients of repeatability for CCT were 0.61, 0.82 and 0.80, whereas mean coefficients of interoperator reproducibility for CCT were 0.91, 1.11, and 1.25.

Conclusions.

CCT and MCT measurements show moderate agreement between instruments.

The repeatability and interoperator reproducibility of the results obtained by SOCT are somewhat higher.

The operator's impact on CCT and MCT measurements is insignificant in all devices.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Piotrowiak, Ilona& Soldanska, Beata& Burduk, Mateusz& Kaluzny, Bartlomiej J.& Kaluzny, Jozef. 2012. Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthalmology،Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-504825

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Piotrowiak, Ilona…[et al.]. Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthalmology No. 2012 (2012), pp.1-5.
https://search.emarefa.net/detail/BIM-504825

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Piotrowiak, Ilona& Soldanska, Beata& Burduk, Mateusz& Kaluzny, Bartlomiej J.& Kaluzny, Jozef. Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthalmology. 2012. Vol. 2012, no. 2012, pp.1-5.
https://search.emarefa.net/detail/BIM-504825

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references

رقم السجل

BIM-504825