Dose Distributions of an 192Ir Brachytherapy Source in Different Media
المؤلفون المشاركون
Liu, Y. W. Hsueh
Hsu, S. M.
Hung, S. K.
Wu, C. H.
Lee, M. S.
Liao, Y. J.
المصدر
العدد
المجلد 2014، العدد 2014 (31 ديسمبر/كانون الأول 2014)، ص ص. 1-11، 11ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2014-04-06
دولة النشر
مصر
عدد الصفحات
11
التخصصات الرئيسية
الملخص EN
This study used MCNPX code to investigate the brachytherapy 192Ir dose distributions in water, bone, and lung tissue and performed radiophotoluminescent glass dosimeter measurements to verify the obtained MCNPX results.
The results showed that the dose-rate constant, radial dose function, and anisotropy function in water were highly consistent with data in the literature.
However, the lung dose near the source would be overestimated by up to 12%, if the lung tissue is assumed to be water, and, hence, if a tumor is located in the lung, the tumor dose will be overestimated, if the material density is not taken into consideration.
In contrast, the lung dose far from the source would be underestimated by up to 30%.
Radial dose functions were found to depend not only on the phantom size but also on the material density.
The phantom size affects the radial dose function in bone more than those in the other tissues.
On the other hand, the anisotropy function in lung tissue was not dependent on the radial distance.
Our simulation results could represent valid clinical reference data and be used to improve the accuracy of the doses delivered during brachytherapy applied to patients with lung cancer.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Wu, C. H.& Liao, Y. J.& Liu, Y. W. Hsueh& Hung, S. K.& Lee, M. S.& Hsu, S. M.. 2014. Dose Distributions of an 192Ir Brachytherapy Source in Different Media. BioMed Research International،Vol. 2014, no. 2014, pp.1-11.
https://search.emarefa.net/detail/BIM-510361
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Wu, C. H.…[et al.]. Dose Distributions of an 192Ir Brachytherapy Source in Different Media. BioMed Research International No. 2014 (2014), pp.1-11.
https://search.emarefa.net/detail/BIM-510361
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Wu, C. H.& Liao, Y. J.& Liu, Y. W. Hsueh& Hung, S. K.& Lee, M. S.& Hsu, S. M.. Dose Distributions of an 192Ir Brachytherapy Source in Different Media. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-11.
https://search.emarefa.net/detail/BIM-510361
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-510361
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر