الملخص EN

A technique has been described to optimize time resolution of NE213 scintilla-tion detectors.

The technique involves minimization of transit time spread and space charge effects in the photomultiplier.

The photomultiplier space charge effect was minimized by optimizing the voltage between the photocathode and the first dynode while the transit time spread was minimized by using a 6th dynode signal as timing signal instead of a conventional anode signal.

During these studies the time resolution of the detector for correlated gamma rays from 22Na and “Co as well as their ratio was used as a check for optimization of detector time resolution.

The detector time resolution is improved by 14 and 21 % for 22Na and 60Co sources when voltage between the photocathode and the first dynode is optimized.

By using the 6th dynode signal for timing analysis, a further improvement of 4 and 6 % is achieved in the detector time resolution for 22Na and 60Co sources respectively.

نمط استشهاد جمعية علماء النفس الأمريكية (APA)

Naqvi, A. A.& Gult, K.. 1991. A technique to optimize the time resolution of NE213 scintillation detectors. The Arabian Journal for Science and Engineering،Vol. 16, no. 4(s), pp.463-470.
https://search.emarefa.net/detail/BIM-598442

نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)

Naqvi, A. A.& Gult, K.. A technique to optimize the time resolution of NE213 scintillation detectors. The Arabian Journal for Science and Engineering Vol. 16, no. 4 (Oct. 1991), pp.463-470.
https://search.emarefa.net/detail/BIM-598442

نمط استشهاد الجمعية الطبية الأمريكية (AMA)

Naqvi, A. A.& Gult, K.. A technique to optimize the time resolution of NE213 scintillation detectors. The Arabian Journal for Science and Engineering. 1991. Vol. 16, no. 4(s), pp.463-470.
https://search.emarefa.net/detail/BIM-598442

نوع البيانات

مقالات

لغة النص

الإنجليزية

الملاحظات

Includes bibliographical references : p. 469-470

رقم السجل

BIM-598442